Fluke 85RF II Скачать руководство пользователя страница 1

 English  

Page 

 

 

 

85RF II 

High Frequency Probe 

Instruction Sheet

 

Introduction  

The 85RF II High Frequency Probe (the Probe) transforms a 
dc voltmeter into a high frequency (100 kHz to 500 MHz) ac 
voltmeter. Conversion from ac to dc is done on a one-to-one 
basis and includes a range of 0.25 V rms to 30 V rms. The 
Probe’s dc output is calibrated to be equal to the rms value 
of a sine wave input. 

Safety 

Symbol Description 

W

 

See Instruction Sheet. 

X

 

Hazardous voltage. 

~

 

Do not dispose of this product as 
unsorted municipal waste. Go to Fluke’s 
website for recycling information. 

WX

Warning 

To prevent possible electrical shock, fire, or personal 
injury: 

 

Examine the case before you use the Product. 
Look for cracks or missing plastic. Carefully 
look at the insulation around the terminals.

 

 

Limit operation to the specified measurement 
category, voltage, or amperage ratings.

 

 

Do not touch voltages >30 V ac rms, 42 V ac 
peak, or 60 V dc. 

 

 

Do not use the Product around explosive gas, 
vapor, or in damp or wet environments.

 

 

Do not use and disable the Product if it is 
damaged. 

 

Use the Product only as specified, or the 
protections supplied by the Product can be 
compromised.

 

 

Do not exceed the Measurement Category (CAT) 
rating of the lowest rated individual component 
of a Product, Probe, or accessory.

 

 

Keep fingers behind the finger guards on the 
Probes. 

 

Do not connect directly to mains. 

PN 4096837 
September 2011, Rev. 1, 12/11 

©2011 Fluke Corporation, Printed in the U.S.A. All product names are 
trademarks of their respective companies. Specifications are subject to change 
without notice. 

®

Содержание 85RF II

Страница 1: ... Product Look for cracks or missing plastic Carefully look at the insulation around the terminals Limit operation to the specified measurement category voltage or amperage ratings Do not touch voltages 30 V ac rms 42 V ac peak or 60 V dc Do not use the Product around explosive gas vapor or in damp or wet environments Do not use and disable the Product if it is damaged Use the Product only as speci...

Страница 2: ...ue of an input and is calibrated to read rms value of a sine wave Voltage range 0 25 V rms to 30V rms Maximum input voltage 30 V rms 200 V dc Input capacitance Approximately 9 pF Temperature range Operating 10 C to 35 C Storage 40 C to 75 C Humidity 90 R H Output connector Fits standard 0 75 inch dual banana connectors Accessory BNC to Probe Adapter P N 574756 Pollution degree 2 Operating altitude...

Страница 3: ...the Probe 2 Touch the Probe tip to the circuit under test Frequencies greater than 100 MHz To measure frequencies greater than 100 MHz 1 Ground the 4 inch clip lead supplied with the Probe 2 Touch the Probe tip to the circuit under test Note Measurements above 100 MHz must be made with minimum cable lengths and a terminated 50 e system One inch or less of non terminated coaxial cable such as a BNC...

Страница 4: ...bration The performance and calibration test is a verification test to make sure that the 85 RF II is functional and correctly calibrated Use the procedure as an acceptance test and or calibration procedure the Probe has one adjustment which is R2 The recommended calibration interval is one year Note If the Probe cannot be calibrated to the specifications in this document the Probe could be damage...

Страница 5: ...e limits disassemble the Probe refer to the Probe Disassembly Section in this Instruction Sheet and adjust R2 for a DMM measurement between 0 990 V dc and 1 010 V dc Reassemble the Probe 3 Move the unit in test to the Normal output of the calibrator and decrease the output level to 0 250 V rms and 1 MHz The DMM will read between 0 223 V dc and 0 280 V dc 4 Set the AC Calibrator to STDBY decrease t...

Страница 6: ...of the printed circuit board up inch and pull it away from the Probe tip 6 To reassemble the Probe do the steps in this procedure backwards start with step five and end with step one Make sure that the gold pin on the front of the printed circuit board aligns correctly in the Probe tip and that the ground lead nut is aligned with the hole in the Probe body How to Clean the Probe WXWarning For safe...

Отзывы: