80TK Calibration Procedure 

A calibration cycle of 1 year is recommended to maintain the 
80TK within specifications. The required equipment is listed 
below. Use Calibration Procedure 1 to calibrate the 80TK for 
use with any K-type thermocouple. Use Calibration Procedure 
2 if the 80TK will be used exclusively with one K-type 
thermocouple. Always calibrate the Celsius function before 
calibrating the Fahrenheit function. 
 

 

 

Location of Calibration Adjustments 

Calibration Procedure 1. 

Use this procedure to calibrate the 80TK: 
1.

 

Verify the condition of the battery as described above and 
replace the battery if necessary. Follow the Battery

 

Replacement procedure to disassemble the 80TK, but do 
not remove the battery or reassemble the 80TK. 

2.

 

Allow 80TK and the room-temperature water bath to 
stabilize at room temperature, away from drafts, for at 
least 30 minutes before proceeding with calibration. Place 
the reference thermometer into the room temperature 
water bath. Allow the reading to stabilize. 

3.

 

Use the shorting bar to short the 80TK thermocouple input 
terminals. 

4.

 

Plug the 80TK into the dc volts input of the multimeter. Set 
the multimeter to the 300 mV or equivalent range. 

5.

 

Set the switch on the 80TK to CELSIUS position. Adjust 
R4 to half rotation, then adjust R5 until the reading on the 
multimeter matches the reading on the mercury reference 
thermometer 

±

 5.0 

°

C. Now adjust R4 until the two 

readings are within 

±

 0.1 

°

C. 

Note 

If the mercury reference thermometer displays °F, 
use the following conversion to get °C: 
[5/9 (temp in °F)-32] = (temp in °C) 

6.

 

Set the switch on the 80TK to the FAHRENHEIT (center) 
position. 

7.

 

Adjust R16 until the reading on the multimeter matches the 
reading on the mercury reference thermometer ± 0.2°F. 

Note 

If the mercury reference thermometer displays °C, 
use the following conversion to get °F:[9/5(temp in 
°C) + 32] = (temp in °F) 

8.

 

Set 80TK switch to the OFF position. Remove 80TK from 
multimeter and remove any connections to the 80TK 
thermocouple input terminals. 

9.

 

Reassemble the 80TK. 

10.

 

The 80TK is now calibrated. 

Calibration Procedure 2. 

Use this procedure if the 80TK will be used exclusively with 
one K-type thermocouple: 
Perform Calibration Procedure 1. Substitute the following for 
step 3 in Calibration Procedure 1: 
3.

 

Connect the thermocouple to the 80TK thermocouple 
input, and place the thermocouple into the room 
temperature bath along with the mercury reference 
thermometer. 

Required Equipment 

Refer to Table 1 for equipment required to calibrate the 80TK. 

Table 1. Required Equipment 

Instrument Minimum 

Specification Recommended 

Model 

Mercury Reference Thermometer 

0.1 

°

C resolution 

Princo Model SAMA-CP45 

Dewar Flask and Cap(water bath) 

2 pint capacity, filled with water 

Thermos bottle 

Digital Voltmeter 

300 mV range, 100 

µ

V resolution 

Fluke Model 87 Series V or equivalent 

Shorting bar  
       OR 
K-type thermocouple 

<1 

  

 
meets NBS standards(with subminiature K plug) 

Wire  
 
Fluke 80PK-1 

 

To contact Fluke or for service, call one of the following telephone numbers: 

USA: 1-888-44-FLUKE (1-888-443-5853) 
Canada: 1-800-36-FLUKE (1-800-363-5853) 
Europe: +31 402-675-200 
Japan: +81-3-3434-0181 
Singapore: +65-738-5655 
Anywhere in the world: +1-425-446-5500 
Or, visit Fluke's Web site at 

www.fluke.com

 

 

 

Fluke Corporation  

Fluke Europe B.V. 

 

 

P.O. Box 9090  

P.O. Box 1186 

 

 

Everett WA 98206-9090 

5602 B.D. Eindhoven 

  

USA 

 

The 

Netherlands 

1.888.475.5235

[email protected]

Fluke

-

Direct

.com

www.valuetronics.com

Содержание 80TK

Страница 1: ...e is suspected of interfering with this module either turn off the transmitter or increase the separation between the transmitter and the module Environmental Ambient Operating Range 0 to 50 C Storage Temperature 40 to 60 C 40 to 140 F Humidity 0 to 90 0 to 35 C 0 to 70 35 to 50 C Basic Accuracy 23 5 C Calibration Applies for one year after purchase or recalibration RANGE ACCURACY of reading degre...

Страница 2: ...illivolts the range has i e 320 C or F if the millivolt range reads 320 mV 3 Use the dc volt setting and ignore the decimal for higher temperatures than can be accommodated by the millivolt range Less resolution is available when using the volt range XWWarning To avoid electrical shock if connecting wires to the input screws make sure the thermocouple is removed from the measurement surface Comple...

Страница 3: ...ter position 7 Adjust R16 until the reading on the multimeter matches the reading on the mercury reference thermometer 0 2 F Note If the mercury reference thermometer displays C use the following conversion to get F 9 5 temp in C 32 temp in F 8 Set 80TK switch to the OFF position Remove 80TK from multimeter and remove any connections to the 80TK thermocouple input terminals 9 Reassemble the 80TK 1...

Страница 4: ...s not apply to fuses disposable batteries or to any product which in Fluke s opinion has been misused altered neglected or damaged by accident or abnormal conditions of operation or handling Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non defective media Fluke does not warrant that so...

Отзывы: