Fluke 80PK-1 Скачать руководство пользователя страница 2

Output: 25

°

C (77

°

F) = 1.00mV (reference junction at 0

°

C)

Seebeck Coefficient: 25

°

C (77

°

F) = 40.50 

µ

V / 

°

C

Measurement Time (Time Constant): 2 seconds (for air at room
temperature and one atmosphere of pressure moving with a
velocity of 65 ft / sec)
Maximum Voltage: 24V ac rms or 60V dc maximum for proper
safe operation.
Maximum Temperature of Bead: 260

°

C (500

°

F

Cable:

Length: 40 inches (1 meter)
Insulation:

Material: Teflon
Maximum Temperature: 260

°

C (500

°

F)

Conductors:

Type K
Size: AWG #24 solid
Length: 48 inches nominal

Polarity and Color Coding:

Chromel: + yellow
Alumel: - red
Outer Jacket: black

Connector:

Type: yellow mini-thermocouple connector with 0.312-inch pin
spacing
Material: Hytrel
Maximum Temperature: 200

°

C (392

°

F)

Protection:

Class 3. Relates solely to insulation and grounding properties
defined in IEC 348.

MEASUREMENT CONSIDERATIONS

Instrument Compatibility

The 80PK-1 is designed to be compatible with any temperature
measuring instrument that accepts Type K thermocouples, has a
miniature thermocouple connector, and has cold reference

1.800.561.8187

[email protected]

Fluke-Direct

.ca