Compatibility
The Product is compatible with any Fluke ScopeMeter test tool, Power
Harmonics Analyzer, Oscilloscope, Multimeter, or other voltage measurement
device with these features:
• BNC input connector (PN PM9081/001 BNC-to-banana adapter) for
use with standard inputs on a digital multimeter (DMM).
• Input accuracy of 2 % or better to take full advantage of the accuracy
of the Probe.
• Input impedance of
≥
1 M
Ω
in parallel with a maximum of 100 pF.
• A passband of more than four times the frequency of the waveform to
be measured.
Instructions
To use the Probe:
1.
Connect the Probe to the input on the measuring instrument. When
the ScopeMeter test tool or an oscilloscope is used, it must have
DC-coupled input. When you are using a digital multimeter, use the
BNC-to- banana adapter (PM9081/001) to connect the Probe to the
input.
2.
On the Probe, select the least sensitive range (10 mV/A). Make sure that
the green ON-indicator lights. See Figure 4 for selector switch and green
On indicator locations.
3.
On the Probe, rotate the ZERO thumbwheel to adjust the reading to
zero. See Figure 4 for the ZERO rotary knob location.
4.
Select the appropriate probe sensitivity on your ScopeMeter test tool or
oscilloscope.
5.
Clamp the Probe around the conductor. Make sure that the arrow mark
on the jaw of the Probe points toward the correct orientation. See
Figure 3.
Figure 3. Current Probe Orientation
6.
Observe the current value or waveform on your display or the current
value readout on the multimeter.
7.
On the ScopeMeter test tool, adjust the vertical range knob and time
division knob for the best display.
8.
If the red OL-indicator lights, the Probe is in overload mode.
80i-110s.book Page 5 Monday, December 5, 2016 12:12 PM