The Product is compatible with any Fluke ScopeMeter test tool, Power 
Harmonics Analyzer, Oscilloscope, Multimeter, or other voltage measurement 
device with these features:
• BNC input connector (PN PM9081/001 BNC-to-banana adapter) for 

use with standard inputs on a digital multimeter (DMM).

• Input accuracy of 2 % or better to take full advantage of the accuracy 

of the Probe.

• Input impedance of 

1 M

 in parallel with a maximum of 100 pF.

• A passband of more than four times the frequency of the waveform to 

be measured. 

Instructions

To use the Probe:
1.

Connect the Probe to the input on the measuring instrument. When 
the ScopeMeter test tool or an oscilloscope is used, it must have 
DC-coupled input. When you are using a digital multimeter, use the 
BNC-to- banana adapter (PM9081/001) to connect the Probe to the 
input.

2.

On the Probe, select the least sensitive range (10 mV/A). Make sure that 
the green ON-indicator lights. See Figure 4 for selector switch and green 
On indicator locations.

3.

On the Probe, rotate the ZERO thumbwheel to adjust the reading to 
zero. See Figure 4 for the ZERO rotary knob location.

4.

Select the appropriate probe sensitivity on your ScopeMeter test tool or 
oscilloscope.

5.

Clamp the Probe around the conductor. Make sure that the arrow mark 
on the jaw of the Probe points toward the correct orientation. See  
Figure 3.

Figure 3. Current Probe Orientation

6.

Observe the current value or waveform on your display or the current 
value readout on the multimeter.

7.

On the ScopeMeter test tool, adjust the vertical range knob and time 
division knob for the best display.

8.

If the red OL-indicator lights, the Probe is in overload mode.

80i-110s.book  Page 5  Monday, December 5, 2016  12:12 PM

Содержание 80i-110s

Страница 1: ...m The 80i 110s is also compatible with any instrument capable of millivolt measurements The Probe provides these benefits Accurate AC DC and AC DC current measurements for Electrical Electronic and Au...

Страница 2: ...read all instructions Do not use the Product if it is altered or damaged Do not alter the Product and use only as specified or the protection supplied by the Product can be compromised Comply with loc...

Страница 3: ...ors is permitted Conforms to European Union directives Conforms to relevant Australian Safety and EMC standards Measurement Category III is applicable to test and measuring circuits connected to the d...

Страница 4: ...he Probe is in the OFF position 3 Locate the battery cover on the handle and loosen the screw with a flat blade screwdriver 4 Slide the battery cover away from the Probe 5 Install the battery IEC 6LR6...

Страница 5: ...NC to banana adapter PM9081 001 to connect the Probe to the input 2 On the Probe select the least sensitive range 10 mV A Make sure that the green ON indicator lights See Figure 4 for selector switch...

Страница 6: ...rying conductors On the Probe the 100 mV A range offers the best accuracy Maintenance Before each use inspect the Probe Look for cracks or missing portions of the housing and output cable insulating c...

Страница 7: ...p cloth and detergent Do not use abrasives or solvents Open the jaws and wipe the magnetic pole pieces with a lightly oiled cloth Do not allow rust or corrosion to form on the magnetic core ends If th...

Страница 8: ...ic purposes There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances Caution This equipment is not intended for us...

Страница 9: ...perature from 0 C to 50 C 32 F to 132 F Performance Verification Verify Probe accuracy with the test setup shown in Figure 5 Required test equipment is defined in Table 1 Toroid coil construction is i...

Страница 10: ...turns wound on cylindrical form using 10 gauge magnet wire Remove form and tape coil together making sure 10 wires are taped Tape D Fan out coil to approximately 270 with spacing 5 cm 2 in using tape...

Страница 11: ...6 A 559 mV 641 mV Table 4 Performance Test Points Current Range 0 A to 100 A 10 mV A DC Measurement 5520A Settings DC Amps Measured Low Limit Output High Limit Output 1 A 10 A 95 5 mV 104 5 mV 3 A 30...

Страница 12: ...uthorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE A...

Отзывы: