44
Diode Test and Continuity Test
Diode test indication:
Displays voltage drop across
device, 2.0 V full scale. Nominal test current 0.2 mA at 0.6
V. Accuracy ±(2 % + 1 count).
Continuity test indication:
Continuous audible tone for
test resistance < 100
Ω
Open circuit voltage
: < 2.9 V
Short circuit current:
220
μ
A typical
Overload protection:
1000 V rms
Loop Power Supply
Voltage:
24 V, Short Circuit protected
DC Current Output
Source mode:
Span: 0 mA or 4 mA to 20 mA, with overrange to 24 mA
Accuracy: 0.05 % of span
1
Compliance voltage: 28 V with battery voltage >~4.5 V
Simulate Mode:
Span: 0 mA or 4 mA to 20 mA, with overrange to 24 mA
Accuracy: 0.05 % of span
1
Loop voltage: 24 V nominal, 48 V maximum, 15 V
minimum
Compliance voltage: 21 V for 24 V supply
Burden voltage: < 3 V
General Specifications
Maximum voltage applied between any jack and earth
ground:
1000 V
Storage temperature:
-40
°
C to 60
°
C
Operating temperature:
-20
°
C to 55
°
C
Operating altitude:
2000 meters maximum
Temperature coefficient:
0.05 x specified accuracy per
°
C for temperatures < 18
°
C or > 28
°
C
1
0.1 x specified
accuracy per
°
C for temperatures < 18
°
C or > 28
°
C