7526A
Calibration Manual
12
Performance Tests
Warning
To prevent possible electrical shock, fire, or personal injury, do
not perform the performance test procedures unless the
Product is fully assembled.
The performance tests verify the complete operation of the Product and measure the
accuracy of each function against Product specifications. If the Product fails a part of the
test, calibration adjustment or repair is necessary. See “Calibration Adjustment”.
Required Test Equipment
The equipment listed in Table 3 is necessary to do the performance tests and calibration
adjustment.
Table 3. Required Equipment
Equipment
Recommended Model
Where Used
8-1/2 digit Precision Digital
Multimeter (DMM)
Fluke 8508A
DC volts, dc current, resistance
Ohms shorting block
Low ohms measure and SPRT
calibration and high ohms
measure calibration
1
Ω
Resistance Standard
Fluke 742A-1
DC current
Resistance standard, 10
Ω
Fluke 742A-10
100 mA source calibration
Resistance standard, 100
Ω
Fluke 742A-100
Low ohms measure and SPRT
calibration
1 K
Ω
Resistance Standard
Fluke 742A-1K
Resistance
10 K
Ω
Resistance Standard
Fluke 742A-10K
Resistance
Multifunction Calibrator
Fluke 5522A
DC volts, dc current, resistance
Thermistor probe
Fluke5640-D
TC CJC calibration
Thermocouple probe
Omega E-type thermocouple
probe
TC CJC calibration
Pressure Module
Any Fluke 700 Series Module
Pressure
Lead Set
5520A-525A/LEADS Leads Kit
Necessary test cables
Precision Thermistor
Fluke 5640-D
Temperature
Thermometer Readout
Fluke 1504
Temperature
Dewar Flask with Lid
- - -
Thermocouple
Characterized Type E
Thermocouple
- - -
Thermocouple
Banana Jack to Copper TC Mini-
connector
- - -
Thermocouple
Lag bath
TC CJC calibration