74X Series
Calibration Manual
6
Frequency Sourcing
Range
Accuracy
0.00 Hz to 10.99 Hz
0.01 Hz
11.00 Hz to 109.99 Hz
0.1 Hz
110.0 Hz to 1099.9 Hz
0.1 Hz
1.100 kHz to 21.999 kHz
0.002 kHz
22.000 kHz to 50.000 kHz
0.005 kHz
Waveform Choices: Zero-symmetric sine or positive square wave, 50% duty cycle
Amplitude: 0.1 to 10V p-p
Amplitude Accuracy:
0 Hz to 1099 Hz: 3% of 0.5% f.s
1.1 kHz to 10.9 kHz: 10% of 0.5% f.s
11 kHz to 50 kHz: 30% of 0.5% f.s
Maximum Input Voltage: 30V dc
Temperature, Thermocouples
Temperature, TCs
Type
Range
°
C
Measure °C
Source °C
1 Year
2 Year
1 Year
2 Year
E
-250 to -200
1.3
2.0
0.6
0.9
-200 to -100
0.5
0.8
0.3
0.4
-100 to -600
0.5
0.8
0.3
0.4
600 to 1000
0.4
0.6
0.2
0.3
N
-200 to -100
1.0
1.5
0.6
0.9
-100 to 900
0.5
0.8
0.5
0.8
900 to 1300
0.6
0.9
0.3
0.4
J
-210 to -100
0.6
0.9
0.3
0.4
-100 to 800
0.3
0.4
0.2
0.3
800 to 1200
0.5
0.8
0.2
0.3
K
-200 to -100
0.7
1.0
0.4
0.6
-100 to 400
0.3
0.4
0.3
0.4
400 to 1200
0.5
0.8
0.3
0.4
1200 to 1372
0.7
1.0
0.3
0.4
T
-250 to -200
1.7
2.5
0.9
1.4
-200 to 0
0.6
0.9
0.4
0.6
0 to 400
0.3
0.4
0.3
0.4
B
600 to 800
1.3
2.0
1.0
1.5
800 to 1000
1.0
1.5
0.8
1.2
1000 to 1820
0.9
1.3
0.8
1.2
Содержание 74X Series
Страница 2: ......
Страница 6: ...74X Series Calibration Manual iv...
Страница 8: ...74X Series Calibration Manual vi...