Automatic Pressure Calibrator
Maintenance
41
Battery Replacement
Replace the battery when it no longer holds a charge for the rated interval. The battery
normally lasts for up to 300 charge/discharge cycles. To order a replacement battery, see
and
User-Replaceable Parts and Accessories
.
Note
Take spent batteries to a qualified recycler or hazardous materials handler for
disposal. Contact an authorized Fluke Service Center for recycling information.
To replace the battery:
1. Push
S
and select Maintenance.
2. Push
A
(
Exhaust
) to release Product internal pressure.
3. Turn the Product Off.
4. Remove test leads.
5. Make sure the Product is unplugged from its charger.
6. Turn the Product over.
7. Unlock the two fasteners to release the battery door. See Figure
8. Remove the battery door.
9. Replace the battery.
10.Replace and lock the battery door.
Figure 17. Battery Replacement