![Fluke 729 Pro Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/fluke/729-pro/729-pro_user-manual_2310413030.webp)
729 Pro
Users Manual
24
Pressure Transmitter (P/P)
To calibrate a pressure-to-pressure transmitter, use this feature. This feature requires a
pressure module to measure pressure output from a connected transmitter.
To use the function, see Figure
1. From page 1 of the Tasks menu, push the arrows to highlight
Pressure Transmitter
(P/P)
.
2. Push
E
or
R
.
3. From this screen, use the arrows, numeric keypad, and
E
to change the values of
the source pressure and measure pressure.
4. Push
D
(
Continue
) and these parameters can be edited:
•
Tolerance of Range
•
Test Strategy
•
Auto Settling Time
5. Push
B
(
Save as Customized
) to save these settings as a customized task.
Recall this customized task from the Tasks menu at any time.
Figure 9. Pressure Transmitter (P/P) Connections
+
+
In Out
Pressure
Module
Pneumatic
Transmitter
Pneumatic
Transmitter
Pressure
Module