Pressure Calibrator
Product Features
17
Example:
A pressure transmitter under test is 30 psi (2 Bar) Full-
Scale and outputs a corresponding 4 mA to 20 mA signal.
Program a 0 psi to 30 psi pressure span into the Product
and the Product calculates and shows the deviation or
%Error from the expected 4 mA to 20 mA output. This
eliminates the need for manual calculations and helps
when it is difficult to set an exact pressure.
To use the %ERROR function:
1.
Push
to access the menus.
2.
Push
to step through the menu until
%ERROR
is
shown on the menu bar.
3.
Push
to open the %ERROR screen.
4.
Push
to scroll through the port choices (P1 or P2).
5.
Push
to configure the pressure range settings.
6.
Use the arrow keys to set the 100 % point of the
desired pressure range, select DONE SET when
finished.
7.
Use the arrow keys to set 0 % point and select
DONE SET. The % ERROR mode is now ready to
use.
Note
The 0 % and 100 % point will be saved in non-
volatile memory until they are changed again by
the user.