![Fluke 714B Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/fluke/714b/714b_user-manual_2310394033.webp)
Thermocouple Calibrator
Specifications
27
Specifications
Specifications are based on a one year calibration cycle
and apply from +18
°
C to +28
°
C unless stated otherwise.
All specifications assume a 5 minute warmup period.
Thermocouple mV Input
Range Resolution
Accuracy (% of R Floor)
1 year
2 year
-10 mV to 75 mV
0.001 mV
0.015 % + 10 µV
0.02 % + 15 µV
Temperature coefficient: ±(0.002 % of r 0.002 % of range) /
°
C ( <18
°
C or >28
°
C)
Thermocouple mV Output
Range Resolution
Accuracy (% of Floor)
1 year
2 year
-10 mV to 75 mV
0.01 mV
0.015 % + 10 µV
0.02 % + 15 µV
Temperature Coefficient: ± (0.002 % of 0.002 % of range) /
°
C ( <18
°
C or >28
°
C)
Thermocouple mA Input
Range Resolution
Accuracy (% of R Floor)
1 year
2 year
0 mA to 24 mA
0.001 mA
0.01 % + 2 µA
0.02 % + 4 µA
Temperature Coefficient: ± (0.002 % of 0.002 % of range) /
°
C ( <18
°
C or >28
°
C)