Before You Start
1.3 Safety Information
1
•
Before initial use, or after transport, or after storage in
humid or semi-humid environments, or anytime the
instrument has not been energized for more than 10 days,
the instrument needs to be energized for a "dry-out" period
of 2 hours before it can be assumed to meet all of the
safety requirements of the IEC 1010-1. If the product is wet
or has been in a wet environment, take necessary
measures to remove moisture prior to applying power such
as storage in a low humidity temperature chamber
operating at 50
°
C for 4 hours or more.
•
Overhead clearance is required. Do not place the
instrument under a cabinet or other structure. Always leave
enough clearance to allow for safe and easy insertion and
removal of probes.
•
The instrument is intended for indoor use only.
•
The bath is a precision instrument. Although it has been
designed for optimum durability and trouble free operation,
it must be handled with care. Position the bath before the
tank is filled with fluid by rolling it into place. DO NOT
attempt to lift the bath. DO NOT move a bath filled with
fluid.
BURN HAZARD
•
Extremely cold temperatures may be present in this
equipment. Freezer burns and frostbite may result if
personnel fail to observe safety precautions.
•
High temperatures may be present in this equipment.
Fires and severe burns may result if personnel fail to
observe safety precautions.
ELECTRICAL HAZARD
•
These guidelines must be followed to ensure that the
safety mechanisms in this instrument will operate
properly. This instrument must be plugged into an AC
electric outlet of the appropriate voltage and frequency
(see Specifications). The power cord of the instrument is
equipped with a three-pronged grounding plug for your
protection against electrical shock hazards. It must be
plugged directly into a properly grounded three-prong
receptacle. The receptacle must be installed in
accordance with local codes and ordinances. Consult a
qualified electrician. DO NOT use an extension cord or
adapter plug.
•
DO use a ground fault interrupt device. This instrument
contains a fluid. A ground fault device is advised in case
fluid is present in the electrical system and could cause
an electrical shock.
Содержание 7108
Страница 6: ...7108 User s Guide iv ...
Страница 8: ...7108 User s Guide vi ...
Страница 10: ...7108 User s Guide viii ...
Страница 18: ...7108 Users Guide 1 8 ...
Страница 24: ...7108 Users Guide 3 4 ...
Страница 31: ...Parts and Controls 6 2 Back Panel 6 6 3 fig_2 eps Figure 2 Back Panel ...
Страница 32: ...7108 Users Guide 6 4 ...
Страница 55: ...Digital Communication Interface 9 3 Interface Commands 9 9 5 Table 5 Interface Command Summary ...
Страница 56: ...7108 Users Guide 9 6 Table 6 Interface Command Summary continued ...
Страница 58: ...7108 Users Guide 10 2 ...