of the fluid and no more than 10 degrees above the upper temperature limit of
the bath.
If the cutout is activated because of excessive bath temperature, power to the
heater shuts off and the bath cools. The bath cools until it reaches a few degrees
below the cutout set-point temperature. At this point the action of the cutout is
determined by the setting of the cutout mode parameter. The cutout has two
selectable modes — automatic reset or manual reset. If the mode is set to auto-
matic, the cutout automatically resets itself when the bath temperature falls be-
low the reset temperature allowing the bath to heat up again. If the mode is set
to manual, the heater remains disabled until the user manually resets the cutout.
The cutout set-point may be accessed within the secondary menu. Press “SET”
and “EXIT” to enter the secondary menu and show the heater power. Then
press “SET” twice to access the cutout set-point.
S+E
Access heater power in secondary menu
12 Pct
Heater power in percent
S
Access proportional band
Pb=0.101C
Proportional band setting
S
Access cutout set-point
CO= 110C
cutout set-point
To change the cutout set-point press “UP” or “DOWN”.
CO= 75C
New cutout set-point
To accept the new cutout set-point press “SET”.
S
Accept cutout set-point
The next function is the configuration menu. Press “EXIT” to resume display-
ing the bath temperature.
9.9
Controller Configuration
The controller has a number of configuration and operating options and calibra-
tion parameters which are programmable via the front panel. These are ac-
cessed from the secondary menu after the cutout set-point function by pressing
45
9 Controller Operation
Controller Configuration