Fluke 700PTP-1 Скачать руководство пользователя страница 2

 

 

Features 

1

2

9

6

8

7

3

5

4

euz01f.eps 

No. Item 

Function 

A

 

Master 
instrument port 

Port for Fluke 700 Series Pressure Module 

B

 

Pressure / 
vacuum button 

Push in from this side for vacuum, and from 
the other side for pressure. Use a small 
screwdriver. 

C

 

Test pressure 
port hose 

Flexible hose to pressure instrument under 
test 

D

 

Quick 
Connects 

1/8 in NPT male 

E

 

Maximum 
stroke limiter 

To limit the pressure from the hand pump, 
tighten these nuts to shorten the compression 
stroke. 

F

 

Pressure 
release valve 

Unscrew to vent pressure from the pump. 
Tighten the valve to close the valve before 
applying pressure. 

G

 

Fine 
adjustment 
knob 

To precisely adjust the applied pressure 

Adapter 

1/8 in NPT female to 1/4 in BSP female 

(2) Clean out 
ports 

See “Pump Valve Assembly Cleaning 
Instructions”   

Figure 1. Features 

Using the Pump 

TC

mA

mA

V

V
RTD

RTD

SOURCE

300V
MAX

30V

MAX

30V

MAX

30V

MAX

MEAS

CAT

  

743

DOCUMENTING PROCESS CALIBRATOR

MEAS

SOURCE

SETUP

mA

V

V

Hz

TC

RTD

7

8

9

4

5

6

1

2

3

0

.

S I G N A L

T E S T

+

Red

Black

ENTER

CLEAR

( ZERO)

RANGE

Pressure

Module

Test Pump

UUT

Calibrator

 

euz02c.eps 

Figure 2. Using the Pump 

W

Warning 

To avoid a violent release of pressure, always 

depressurize the system slowly using the pressure 

release valve (item 

F

) before detaching any 

pressure line from the pump. Do not connect the 

pump to an external pressure source. 

1.

 

Attach a 700 Series Pressure Module (hereafter called “pressure 
module”) to the master instrument port. Use Teflon tape or other 
sealing medium on all NPT thread connections to eliminate leakage. 

2.

 

Connect the test pressure port hose (item 

C

) to the pressure input of 

the pressure instrument under test. 

3.

 

Make sure the pressure/vacuum button (item 

B

) is in the desired 

position. 

4.

 

Unscrew the pressure release valve (item 

F

) to vent pressure from 

the pump. 

5.

 

Zero the pressure module. The zeroing procedure depends on which 
pressure module and calibrator is in use. 

6.

 

Turn the fine adjustment knob (item 

G

) to mid-range. 

7.

 

Tighten the pressure release valve (item 

F

). Do not over tighten. 

8.

 

To limit the maximum pressure that can be developed by the pump, 
tighten the maximum-stroke limiting adjustment nuts (item 

E

). This 

also reduces the amount of pressure increase per stroke. 

9.

 

To apply very low pressures, use the fine adjustment knob only (item 

G

). 

10.

 

Squeeze the handles together to apply incrementally higher pressure. 
For pressure over 7 bar (100 psi), use two hands. 

11.

 

Adjust the pressure using the fine adjustment knob (item 

G

). 

Отзывы: