700G Series
Users Manual
10
each parameter. To move to the next parameter, push
.
You can access, lock, or unlock these functions:
•
Zero function (enable/disable)
•
Set pressure units (enable/disable)
•
Auto shutdown adjustment (enable/disable)
•
Damping settings (enable/disable)
•
Sample rate setting (enable/disable)
•
Tare setting (enable/disable)
•
Custom engineering units (set scale factor)
When a function is locked, it cannot be accessed or
changed from its current condition until you go to
Supervisory Mode and unlock it.
Available Pressure Ranges
Available pressure ranges are listed in the Specifications
section.
How to Set a Custom Engineering Unit or Scale
The last menu selection in Supervisory mode is
SET FACTR
. You can set a multiplier factor from 0.001 to
100 to make a custom scale. The set factor is multiplied
by the psi measured and the result is shown.
Example: 40 psi is the equivalent of 1000 lbs of product in
a tank. You want to show the product weight with a 100
psi gauge. If you set a factor of 25, 40 psi pressure would
show as 1000 (40 x 25). The engineering unit shown is
Cust
(custom).
Battery Life
Battery life is approximately 1500 hours (60 days) of
continuous operation with the backlight off. With
intermittent operation, batteries could last a year or more.
When the battery voltage is low, the low-battery icon (
)
shows on the top left of the display. To replace the
batteries, see the How to Change the Batteries section.
Содержание 700G01
Страница 5: ...iii List of Tables Table Title Page 1 Symbols 4 2 Display and Buttons 5...
Страница 6: ...700G Series Users Manual iv...
Страница 7: ...v List of Figures Figure Title Page 1 The Product 5 2 Change the Batteries 11...
Страница 8: ...700G Series Users Manual vi...
Страница 28: ...700G Series Users Manual 20...