67 MAX
Calibration Manual
8
Figure 2. Precision Black Body Source System (PBBS)
Prepare the Black Body Source
Make sure that the black body is powered on and is allowed to warm up for at least 1 hour.
To configure the black body:
1. Remove the target cover.
2. Set the apparent emissivity setting to 0.98.
84.9814 C
.
1
2
3