Infrared Thermometers
Menu Overview
7
•
To enter an emissivity value manually:
1.
Press the
No
. softkey.
2.
Use the down or up arrow softkey to change the entry. Hold down the up or
down arrow softkey to increase the rate of change.
3.
Press the
Done
softkey when finished to return to the main menu.
Table 3. Nominal Surface Emissivity
Material
Value Material Value
Default****
0.95 Glass
(plate)
0.85
Aluminum*
0.30 Iron*
0.70
Asbestos 0.95
Lead*
0.50
Asphalt 0.95
Oil
0.94
Brass*
0.50
Paint
0.93
Ceramic 0.95
Plastic**
0.95
Concrete 0.95
Rubber 0.95
Copper*
0.60 Sand
0.90
Food - Frozen
0.90
Steel*
0.80
Food - Hot
0.93
Water
0.93
Wood ***
0.94
* Oxidized
** Opaque, over 20 mils
*** Natural
**** Factory Setting
Highlighted items may also be found in the emissivity table built into the thermometer.
°
C and
°
F
To toggle between
°
C (Celsius) and
°
F (Fahrenheit) measurements, press the
Menu
softkey until
°
C
or
°
F
appears
as the left softkey function. Press the corresponding
softkey to change between the measurements.
Min, Max, Avg, Differential
The thermometers can measure minimum (MIN), maximum (MAX), average (AVG), or
differential (
Δ
) temperatures each time a reading is taken. These values are not shown if a
thermocouple is plugged into the thermometer.
To turn on the Min Max mode:
1.
Press the
Menu
softkey until
MnMx
(Min Max) appears as the left softkey function.
2.
Press the
MnMx
softkey.
The display shows the present reading, maximum and minimum readings, and the
emissivity setting.
To turn on the Avg/Dif mode:
1.
Press the
Menu
softkey until
Avg
appears as the right softkey function.