Installation

Lead Wire Identification

The 5618B PRT is equipped with a four-wire cable (see Figure 1). Four lead wires are used to 

cancel lead wire resistance. For best results, the readout device should be equipped to handle 

four-terminal resistors. The lead wires are two different colors. Lead wire pairs attached to each 

end of the sensor are identified by red and white insulation.

White

Red

White

Red

5618B Lead Wires.eps

Figure 1. PRT Schematic

PRT Care and Handling Guidelines

Caution  

W

Read this section before removing the PRT from the shipping 

box or case.

Care must be taken in handling the PRT to maintain calibration accuracy. Care should still be used 

when handling the PRT even though the Inconel sheath is durable and provides good protection 

for the sensor. Correct handling of the PRT will prolong the life expectancy. When not in use, the 

PRT should be stored in an optional protective case that can be purchased by contacting Fluke 

(see “How to Contact Fluke”). The handle is not designed to be immersed.

PRT Handling Guidelines

• 

Keep the thermometer as clean as possible.

• 

Immerse the thermometer in the appropriate liquid for the temperature range. If a dry block 

is used, the well diameter should allow the PRT to comfortably slip in and out without excess 

movement. For best results, immerse the thermometer as deep as possible to avoid “stem 

effect” (the temperature error caused by the conduction of heat away from the sensor).

• 

Allow sufficient time for the thermometer to stabilize before making measurements. This 

allows for the best accuracy.

• 

Use the correct drive current with the thermometer to prevent error in temperature or 

resistance. Fluke recommends 1 mA.

• 

Use the protective shipping box provided or other protection when the thermometer is not in 

use.

• 

DO NOT

 subject the thermometer to any physical shock or vibration.

• 

DO NOT

 use pliers or other devices to squeeze the sheath. This action can permanently 

damage the PRT.

• 

DO NOT

 subject the thermometer to temperatures above the highest specified operating 

temperature.

• 

DO NOT

 expose the thermometer’s handle or cables to extreme temperatures.

• 

DO NOT

 screw a clamp down so tight that it dents the sheath. This can permanently damage 

the PRT.

Содержание 5618B

Страница 1: ...Diameter Industrial RTD 2004 Rev 1 4 11 2004 2011 Fluke Corporation All rights reserved Specifications subject to change without notice Users Guide All product names are trademarks of their respectiv...

Страница 2: ...arest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB Des...

Страница 3: ...on 1 Before You Start 1 Symbols Used 1 Safety Information 2 How to Contact Fluke 2 Specifications 3 Installation 4 Lead Wire Identification 4 PRT Care and Handling Guidelines 4 Operation 5 Immersion R...

Страница 4: ...5618B Users Guide ii...

Страница 5: ...iii List of Tables Table Title Table 1 International Electrical Symbols 1 Table 2 Specifications 3 Table 3 PRT Troubleshooting 6 Page List of Figures Figure Title Page Figure 1 PRT Schematic 4...

Страница 6: ...5618B Users Guide iv...

Страница 7: ...t should be checked periodically at the Triple Point of Water TPW For information on drift with respect to mechanical shock and oxidation refer to the Troubleshooting section of this manual Before You...

Страница 8: ...sed DO NOT use the Product around explosive gas vapor or in damp or wet environments Caution To avoid possible damage to the instrument follow these guidelines DO NOT drop or hit the probe in any way...

Страница 9: ...Sheath Dimensions Length x Diameter 5618B 6 152 mm x 3 2 mm 6 in x 0 125 in 5618B 9 229 mm x 3 2 mm 9 in x 0 125 in 5618B 12 305 mm x 3 2 mm 12 x 0 125 in Sheath Material 316 SST Transition Junction...

Страница 10: ...signed to be immersed PRT Handling Guidelines Keep the thermometer as clean as possible Immerse the thermometer in the appropriate liquid for the temperature range If a dry block is used the well diam...

Страница 11: ...annealing the full length of wire before construction of the PRT Likewise connection to extension lead wires and readout instruments can be a source of thermal EMF The thermal EMF is caused by a diffe...

Страница 12: ...or assistance see How to Contact Fluke If the data is unstable at high temperatures it may be due to electrical noise in the system Reduce the temperature and observe the data If it is stable electric...

Отзывы: