50, 1, 2, and 10-Turn Current Coil

Specifications

3

Specifications

Calculate Total Specification

The total specification of the effective current that the clamp meter measures is a function of 

the clamp/coil interaction and of the current calibrator. To determine the total specification 

use this formula:

Example:

A Fluke Calibration 5560A Calibrator supplies 13 A at 60 Hz to the 55XXA/COIL 50 generating 

an effective current of 650 A (50 x 13 A), 60 Hz). The 1-year specification of the Calibrator at 

13 A is ±(0.07 % + 8 mA), which is equivalent to 0.13 % of the 13 A calibrator output. The total 

specification of the calibrator and the coil together as a percentage of the output:

Specification of calibrator current = ±0.13 %
Specification due to the Coil = 0.65 % 

The RSS of these two specifications determines the total specification of the source/coil 

combination:

For higher frequencies or higher current levels, the compliance voltage may exceed 1 V. If the 

compliance voltage exceeds 1 V, then compliance adders must be added to the Calibrator 

specification.

55XXA/COIL 50

55XXA/COIL 10

Number of turns

50

1, 2, or 10

Maximum Current

20 A rms, continuous

30.2 A rms, 5 minutes

30.2 A rms, continuous

Maximum Duty Cycle Derating

<20 A, continuous

>20 A, 50 % Duty Cycle

-

Maximum Voltage, All Inputs

30 V to Earth

Minimum Inner Diameter of Clamp Jaws

23 mm

1-Turn:  3 mm
2-Turn:  6 mm
10-Turn: 10 mm

Coil Ratio

±

0.65 %

[1] 

DC 

to 440 Hz. 

[1] Includes coil/clamp interaction but does not include current source errors.

—

S

2

coil

+ U

2

source

S

total

 =

—

0.112 + 0.562 + 0.66%

S

total

 =

Содержание 55XXA/COIL 50

Страница 1: ...r of 50 to support the calibration and verification of these clamp meters with the 50 turns of the 55XXA COIL 50 Contact Fluke Fluke Corporation operates worldwide For local contact information go to our website www flukecal com To register your product or to view print or download the latest manual or manual supplement go to our website 1 425 446 5500 fluke info fluke com Fluke Corporation P O Bo...

Страница 2: ...e Product if it is altered or damaged Symbols Table 1 lists the symbols used on the Product or in this document Table 1 Symbols Symbol Description Consult user documentation W WARNING RISK OF DANGER X WARNING HAZARDOUS VOLTAGE Risk of electric shock P Conforms to European directives This product complies with the WEEE Directive marking requirements The affixed label indicates that you must not dis...

Страница 3: ... output Specification of calibrator current 0 13 Specification due to the Coil 0 65 The RSS of these two specifications determines the total specification of the source coil combination For higher frequencies or higher current levels the compliance voltage may exceed 1 V If the compliance voltage exceeds 1 V then compliance adders must be added to the Calibrator specification 55XXA COIL 50 55XXA C...

Страница 4: ...l see Figure 1 and Figure 2 55XXA COIL 10 see Figure 1 For 10 turns center the clamp meter on the Coil base For 1 turn and 2 turns align the clamp meter with the markings on the Coil base 55XXA COIL 50 see Figure 2 Center the clamp meter on the base Calibration accuracy to specifications is guaranteed only with proper clamp alignment Align the clamp meter as much as possible on the base during cal...

Страница 5: ...50 1 2 and 10 Turn Current Coil Use the Coil 5 Figure 2 55XXA COIL 50 CO MM ON 10 TU RN 30V ...

Страница 6: ...luke authorized service center to obtain return authorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIED FLUKE IS NOT LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES ARISING FROM ANY CAUSE OR THEORY Sin...

Отзывы: