Calibrator
Menus
19
Setup Menu > Instrument Setup
The
Instrument Setup
menu (
Setup Menu > Instrument Setup
) is the second choice of the list of
submenus shown on the left of the Setup Menu. Parameters are nonvolatile and remain set after you
reset the Product or when the Product power is turned on.
Note
There is a scrollbar on the right of the screen. Place your finger anywhere on the active section
of the display and drag your finger to move the screen content.
The Instrument Setup menu includes:
• Output Limits
•
Set or view the upper and lower ac and dc voltage and terminal current limits.
•
Restore Limit Defaults
• Defaults
•
Set or view the Product defaults
The Product Setup Choices are in Table
.
Table 7. Product Setup Choices
Parameter
Setting Choices
Thermocouple Type
[1]
A1 (BP, A), B, C, D, E, G, J, K, L, N, R, S, T, U, XK, J, N,
T, 10
μ
V/
°
C, 1mV/
°
C
RTD Type
[1]
Pt 100 (3926), Pt 100 (3916), Pt 100 (385), Pt 200 (385),
Pt 500 (385), Pt 1000 (385),
Ni120 (672), Cu10 (427), Cu50 (428), Cu100 (428)
Temperature Unit
[1]
°
C/
°
F
Temperature Scale
[1]
ITS-90, IPTS-68
dBm Reference
[1]
50,100, 300,1k(dBv), 75,135, 600,1200, 90,150, 900
Verification Interval
90 Day, 1 Year, 2 Year
Confidence Level
95%, 99%
Error Reference
Nominal Value, True Value
Reference Clock
Internal, External
Multi-Unit Phase Synchronization Button
Show, Hide
Error Unit
Scientific Notation, Percent, Parts Per Million (x 10-6):
when <10 PPM, <100 PPM, <1000 PPM
Reference Phase Default
[1]
-180.0 to 180.0
Scope Overload Default
1.0 to 60.0 seconds
Show Secondary Specifications
Show or Hide
Restore Factory Defaults
-
[1] Changes to these default values does not affect the current active setting until you Reset, Power Cycle, or re-select the
function through the function menu.