5320A-LOAD 

High Voltage Load Adapter 

Instruction Sheet 

Introduction 

The 5320A-LOAD High Voltage Load Adapter 

(hereafter the Load) is designed for loading a 

hipot tester to create leakage current while 

performing hipot tester calibration with the 5320A 

Electrical Safety Tester Calibrator (hereafter the 

Calibrator). As shown in Figure 1, this Load 

consists of a series of eight power-rated resistors 

configured to provide eight resistance taps from 

10 k

Ω

 to 5 M

Ω

. The Load is designed to 

withstand a maximum voltage of 1.2 kV to 

5.5 kV, depending upon the selected resistor. 

COM

10 k

10 k

20 k

15 k

50 k

150 k

250 k

500 k

4 M

35 k

50 k

100 k

250 k

500 k

1 M

5 M

 

ehq059 

Figure 1. Load Schematic 

XW

Warning 

To avoid possible electric shock or 

personal injury, use this Load only as 

specified in this Instruction Sheet or 

the 5320A Calibrator Users Manual. 

Preparing for Operation 

The Load is shipped with a power line adapter 

(to run the cooling fans), a grounding cable, and 

this Instruction Sheet. 
The power line adapter comes with five different 

mains plug adapters for compatibility with 

different power plug configurations. Prior to use, 

select the appropriate plug adapter which 

matches the local power mains outlet. After 

attaching the plug to the power line adapter it is 

ready for use with the Load. 
After placing the Load on top of the Calibrator, 

connect the grounding cable between the Load’s 

rear-panel ground terminal and the Calibrator’s 

rear-panel ground terminal. 

XW

 Warning 

To avoid electric shock from high-

voltage energy, ensure the ground 

terminal on the rear of the Load is 

connected to the ground (GND) 

terminal on the rear of the Calibrator. 

Any application with non-grounded 

housing is strictly prohibited. 

W

 Caution 

To prevent damage to the Load, 

ensure the vents on the bottom of the 

Load and the fan exhaust vents are 

clear of obstructions for proper 

cooling. 

Connect one end of the power line adapter to the 

mating power input connector socket on the rear 

panel of the Load and the other end into the 

mains power outlet. 

W

Caution 

Failure to have the fans operating 

may lead to over heating and 

component failure. 

 

 

5320A-LOAD

 

1200 V MAX

140 W MAX

2000 V MAX
110 W MAX

2000 V MAX

80 W  MAX

5500 V MAX
300 W MAX

5500 V MAX
120 W MAX

5500 V MAX

60 W MAX

5500 V MAX

30 W MAX

5500 V MAX

6 W MAX

COM

10 k

35 k

500 k

1M

50 k

100 k

250 k

5 M

100 mA

40 mA

10 mA

4 mA

1mA

POWER

I

O

20 V MAX

 

ehq102.eps 

Figure 2. 5320A-LOAD Front Panel