background image

Theory of Operation

Circuit Analysis

3

3-7

the rate at which the input voltage is applied to the A5 Preamplifier when the 5220A is
switched from standby to operate. This is necessary to prevent a step output which, if
applied to an inductive load, would result in an overvoltage transient, forcing the 5220A
back to standby. In standby both input select relays, K1 and K2, are open, capacitors, C3
and C12, are discharged, and the photo resistor U3 is a very high impedance with the
LED drive removed. Thus, the A5 Preamplifier is effectively isolated from the input
voltage.

In the operate mode FET Drive 1 is stepped to –15 V dc causing FET Q1 to turn off.
With Q1 off, the A5 Preamplifier returns to its high gain state with feedback established
by resistors R1 and R2. If front input is selected, the Relay 1 drive signal is low and
capacitor C12 begins charging. The increasing voltage across C12 increased the LED
output from U3 which, in turn, decreases the impedance of the photo resistor. Since the
resistor is in parallel with the input contacts of the still open front input relay K1, it acts
as a variable input attenuator. In other words, the front input voltage is applied to the
input of the A5 Preamplifier at a controlled rate and not as a step input. At the same time,
a low Relay 1 drive signal is inverted by U4 causing C3 to charge through R24. When the
charge reaches the threshold of the second inverter (U4) relay K1 is energized by a set of
four parallel connected inverters. The charge time for C3 and C12 are designed to ensure
that K1 closes after photo resister U3 changes to a low impedance. When K1 closes, it
places a short across the resistor in U3 and selects the front panel input voltage to drive
the A5 Preamplifier. If the rear panel input signal is selected, a variable attenuator is not
necessary because the signal from the remote 5100A rises at a controlled rate when
switched from standby to operate.

3-9. 

A6 Driver

The A6 Diver, as shown in Figure 3-5, is a differential amplifier designed to accept its
input from the A5 Preamplifier and to provide a differential drive signal to the A7 Output
stage. The amplifier itself is of a discrete-component, complementary, push-pull design.
Its features include the following:

 

Temperature compensated biasing for both itself and the A7 Output stage

 

An amplifier-disable circuit for standby operation

 

An output-leakage current sink for standby operation.

The input section of the driver amplifier consists of a differential transistor pair, Q2 and
Q3; a current regulator, Q4; and a common base driver, Q1, for the bias circuit.
Transistor Q3 of the input pair receives the drive input from the A5 Preamplifier. The
other input transistor, Q2, is driven by Shunt Sense 2, which provides negative feedback
from the output current shunt to reduce distortion in the A7 Output stage. Current
regulator Q4 regulates the emitter current of the differential pair, and provides isolation
from variations in the –12 V supply. Output from the differential pair is taken at the
collector of Q3 by way of a common-base driver, Q1.

The bias section of the driver amplifier consists of resistors, R13 through R17, current
regulator, CR1, and transistors, Q5 and Q107 (Q107 is physically located on the output
amplifier's heat sink). Resistors R13 through R17 and current regulator CR1 form a series
string that is driven by the output driver Q1 of the input amplifier. Regulator CR1 is at
the end of the string and connected to –12 V dc. This ensures a constant current of
approximately 5.2 mA through the string. Transistors Q5 and Q107 are connected across
resistors R14, R15 and R16, R17 as VBE multipliers to ensure that the voltage drop
across the resistor string changes at a rate which will compensate for VBE temperature
changes in the A6 Driver and A7 Output stages. Transistor Q5 senses ambient
temperature changes while Q107 senses temperature changes due to power dissipation in
the A7 output. The temperature compensated voltage drop developed across the resistor

Содержание 5220A

Страница 1: ...220A Transconductance Amplifier Instruction Manual PN 491936 June 1979 Rev 3 6 97 1997 Fluke Corporation All rights reserved Printed in USA All product names are trademarks of their respective compani...

Страница 2: ...nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty postage and insurance prepaid FOB D...

Страница 3: ...2 2 9 Operating Notes 2 5 2 10 AC Line Connection 2 5 2 11 Transconductance Amplifier 2 5 2 12 Input Voltage Requirements 2 5 2 13 Input Voltage Selection Front Rear 2 6 2 14 Current Output Front Rea...

Страница 4: ...t Assembly Removal 4 5 4 12 A3 Cap Bus Removal 4 5 4 13 A12 Shunt Assembly Removal 4 5 4 14 Fuse Replacement 4 6 4 15 Ac Line Fuse 4 6 4 16 Power Supply Fuses 4 6 4 17 Line Voltage Selection 4 6 4 18...

Страница 5: ...2 How to Obtain Parts 5 1 5 3 Manual Status Information 5 2 5 4 Newer Instruments 5 2 5 5 Service Centers 5 2 6 Option and Accessory Information 6 1 6 1 Introduction 6 1 6 2 Accessories 6 1 6 3 Rack M...

Страница 6: ...5220A Instruction Manual iv...

Страница 7: ...ing Guide 4 21 4 7 Analog Section Troubleshooting Guide 4 25 5 1 5220A Final Assembly 5 4 5 2 A1 Motherboard PCB Assembly 5 17 5 3 A2 Power Transformer Assembly 5 18 5 4 A3 Capacitor Bus PCB Assembly...

Страница 8: ...5220A Instruction Manual vi...

Страница 9: ...nce 3 14 3 10 A11 MIS Bus Interface 3 18 4 1 Internal Component Assembly Locations 4 3 4 2 Line Voltage Selection 4 7 4 3 Test Point Location Identification 4 19 4 4 Flowchart Summary of Mainframe and...

Страница 10: ...Diagram 7 2 7 2 Chassis Wiring 7 4 7 3 Power Supply 7 6 7 4 A5 Preamplifier PCB Assembly 7 8 7 5 A6 Driver and A7 Output PCB Assemblies 7 10 7 6 A8 Analog Control PCB Assembly 7 12 7 7 A9 Front Panel...

Страница 11: ...Contents continued ix...

Страница 12: ...5220A Instruction Manual x ajs36f eps 5220A Transconductance Amplifier...

Страница 13: ...on A series of front panel indicators are used to visually display both the control and operating status of the 5220A Control indicators include OPER operate STDBY standby FRONT input REAR input REMOT...

Страница 14: ...20 C to 30 C and a relative humidity of 70 or less Basic Transconductance 1 Siemen 1 ampere per volt Output Range 0 A to 20 A dc or rms ac 28 3 A peak Compliance Voltage 4 V dc or 3 V rms ac 4 25 V p...

Страница 15: ...5100 Series B Output Range 1 A rms to 19 9999 A rms Accuracy of Output 0 07 of selected output 1 mA rms from 50 Hz to 1 kHz and 0 07 of selected output 1 mA rms x f from 1 kHz to 5 kHz where f freque...

Страница 16: ...Range Operating 0 C to 50 C Storage 20 C to 56 C Relative Humidity 50 to 50 C 75 to 40 C 95 to 25 C Altitude Operating 0 ft to10 000 ft Non operating 0 ft to 40 000 ft Vibration 2 g maximum 5 Hz to 55...

Страница 17: ...Introduction and Specifications Specifications 1 1 5 ajs01f wmf Figure1 1 Outline Drawings...

Страница 18: ...5 220A Instruction Manual 1 6...

Страница 19: ...er Upon receipt of the instrument a thorough inspection should be made to reveal any possible shipping damage Special instructions for inspection and claims are included on the shipping carton If resh...

Страница 20: ...power supplies from damage in the event of an overload These fuses six in all are internal to the 5220A and are not available to the operator Details for replacement by qualified personnel are given i...

Страница 21: ...Operating Instructions Operating Features 2 2 3 ajs02f wmf Figure 2 1 Controls Indicators and Connectors...

Страница 22: ...remote is enabled and the LOCAL LOCKOUT LED is not lit OPR STBY LEDs One of the two is lit to indicate the current operating mode operate or standby FRONT REAR LEDs One is lit to indicate the selected...

Страница 23: ...greater than the load resistance Similarly its input impedance is much greater than that of the input voltage source The input output ratio of the 5220A is 1 1 and the output current range is 20 to 20...

Страница 24: ...counteracted under certain conditions For example either mode may be remotely selected and the detection of an output overload will command the standby mode Similarly when the unit is initially energi...

Страница 25: ...nt This is illustrated in Figure 2 2 for maximum load inductance vs frequency at output currents of 5 10 and 20 A 2 A step change in output current into an inductive load can easily produce an overcom...

Страница 26: ...5 220A Instruction Manual 2 8 ajs03f wmf Figure 2 2 Maximum Load Inductance vs Frequency...

Страница 27: ...a the Y5000 Interface if remote operation is desired See the 5100B Y5000 Instruction Manuals for details 2 Connect the 5220A to line power 3 Turn on the 5220A by depressing the POWER switch to the ON...

Страница 28: ...e the OPR LED will light indicating that the front CURRENT OUTPUT terminals are active 8 Press the STDBY switch and then remove the short from the front output terminals 9 Connect the intended load be...

Страница 29: ...can be divided into two sections analog and digital The analog section consists of the power supplies the transconductance amplifier and associated analog control circuitry The digital section compri...

Страница 30: ...5 220A Instruction Manual 3 2 ajs05f wmf Figure 3 1 5220A Functional Block Diagram...

Страница 31: ...s sent to the A5 Preamp as a feedback signal thus closing the loop Overall operation of the transconductance amplifier is illustrated in Figure 3 2 When an input voltage EIN is applied to the amplifie...

Страница 32: ...ee of the four are regulated The three regulated voltage sets are 15 15 V dc 5 V dc and 5 V ISO isolated The unregulated voltage set is 12 12 V dc The A2 Transformer assembly consists of a series of t...

Страница 33: ...ating techniques That is it is rectified filtered regulated by a three terminal regulator and filtered again before being output A fifth voltage Line Mon monitor is also produced on the A4 Regulator I...

Страница 34: ...lds the collectors of Q2 at a low and equal level to maintain low and equal collector to base leakage currents ICER at Q2 The actual collector voltage of Q2 is set by the temperature corrected divider...

Страница 35: ...om standby to operate 3 9 A6 Driver The A6 Diver as shown in Figure 3 5 is a differential amplifier designed to accept its input from the A5 Preamplifier and to provide a differential drive signal to...

Страница 36: ...ed in the stage by the 10 ohms resistors R30 and R31 in the collector circuits The differential output at the emitters of Q10 and Q12 is used as the drive input to the A7 output stage In the standby m...

Страница 37: ...ransient or fault conditions Emitter resistors in the parallel output pairs prevent current hogging which would otherwise occur due to unequal VBE characteristics ajs11f wmf Figure 3 6 A7 Output Funct...

Страница 38: ...o standby Each of the detectors is basically the same They consist of an absolute value amplifier full wave rectifier a level conditioner a zener diode threshold detector and a TTL buffered output sta...

Страница 39: ...pression represents a 5 V output to the A10 Logic A series resistor connected to the anode end of each LED serves as a current limiter when the LED is lit A low logic level from the A10 Logic lights t...

Страница 40: ...y to go low Normally the line monitor input is a clipped full wave rectified 60 Hz waveform which is generated on the A4 Regulator Assembly The negative going edge is used as the trigger for monostabl...

Страница 41: ...F LED off Interface Overload Status Low None 3 15 Timing Logic The timing logic controls the timing sequence used to switch the transconductance amplifier between the standby and operate modes It cons...

Страница 42: ...state of the timing logic shift left standby mode or shift right operate Two it drives the front panel STANDBY OPERATE LEDs and provides the A11 MIS Bus Interface remote interface with an Interface O...

Страница 43: ...m monostable U8 9 This pulse clears flip flop U7 which in turn commands the A10 Logic to the local mode Rem Stat low As a result the Remote LED and LLO LED signals are high Interface Remote Status is...

Страница 44: ...ance amplifier They are 1 Thermal Sw itch 1 2 Overcompliance 3 Overcurrent 3 20 A11 MIS Bus Interface The A11 MIS Bus Interface as shown in Figure 3 10 is a remote control interface that allows the 52...

Страница 45: ...laces the correct address on the address bus BC0 and BC2 high the read status logic responds by pulling the not ACK line low and gating the three status bits onto the data bus The not ACK signal is se...

Страница 46: ...5 220A Instruction Manual 3 18...

Страница 47: ...received and later as a calibration procedure to verify the 180 day specifications Table 4 1 lists the equipment required for both the performance test and calibration adjustments If a recommended mod...

Страница 48: ...Filter Maintenance Periodically inspect the air filter on the rear panel for dirt and contaminants If cleaning is required use the following procedure 1 Disconnect the 5220A from line power 2 Remove t...

Страница 49: ...4 4 3 ajs15f wmf Figure 4 1 Internal Component Assembly Locations Caution To prevent component damage observe static awareness precautions during instrument disassembly assembly Refer to the yellow i...

Страница 50: ...Front Panel PCB Assembly 1 Remove the front panel 2 Remove the two screws from the top of the A9 Front Panel PCB Assembly 3 Pull the PCB from its connector J10 4 10 A2 Transformer Assembly Removal Us...

Страница 51: ...d 7 Remove the nine screws holding the A3 Cap Bus to the front panel bulkhead 8 On the A3 Cap Bus locate and the remove the two large head screws located to the side of resistors R2 and R5 Caution Whe...

Страница 52: ...and large inner covers from the 5220A 3 With reference to Figure 4 1 locate and inspect fuses F2 through F7 The fuses are installed in fuse clips and can be pulled out using your fingers 4 Replace de...

Страница 53: ...e current output connections The active terminals are determined by the position of a pair of screws that serve as output jumpers See Figure 4 1 for the output jumper screw locations Installing the sc...

Страница 54: ...libration periods It is also useful as an aid in troubleshooting Test equipment required for the performance test is listed earlier in Table 4 1 If the recommended equipment is not available comparabl...

Страница 55: ...aution The THERMAL CUTOFF LED will light only when the output stage of the 5220A is overheated This is a failure mode that should not be invoked for test purposes The LOCAL pushbutton and the REMOTE L...

Страница 56: ...n steps 2 and 3 of Table 4 2 At each voltage setting verify the 5220A output accuracy by repeating step 7 of this procedure 10 Set the 5220A to the standby mode 11 Set the dc voltage calibrator output...

Страница 57: ...0000 V 100 Hz 4 Set the 5220A to the operate mode 5 Read the true rms voltmeter and calculate the current It should indicate a current within the limits shown for step 1 6 Sequentially set the calibr...

Страница 58: ...A 6 Remove the 1 ohm 10 W resistor from the 5220A output terminals 7 Set the 5220A POWER switch to OFF 4 29 Calibration Adjustments The calibration adjustment procedures given in the following paragra...

Страница 59: ...onnect the DVM to the sense terminals on the load resistor and select the 100 mV dc range 3 Connect a short jumper wire between the front VOLTAGE INPUT terminals 4 Set the 5220A to the operate mode 5...

Страница 60: ...Connect the output of the dc voltage calibrator to the VOLTAGE INPUT terminals on the 5220A positive to red negative to black 4 Set the dc voltage calibrator for an output of 10 0000 V dc 5 Set the 52...

Страница 61: ...bled output terminals on the 5220A 2 Set the ac voltage calibrator output for 3 3 V rms 50 Hz 3 Set the 5220A to the operate mode If the unit switches back to standby turn R9 on the A8 Analog Control...

Страница 62: ...g the STDBY OPR switch 10 Note the DVM reading It should read within the range of 1 300 mV dc 11 Adjust R6 on the A5 Preamplifier PCB to the full cw clockwise position and note the reading The DVM rea...

Страница 63: ...o the display 001 mV 5 Connect the output of the dc voltage calibrator to the VOLTAGE INPUT terminals on the 5220A positive to red negative to black 6 Adjust R3 on the A5 Preamplifier to its full cw l...

Страница 64: ...tion Do not handle ICs or PCBs by their connectors Use static ground straps to discharge repair personnel Use conductive foam to store replacement or removed ICs Remove all plastic vinyl and Styrofoam...

Страница 65: ...Maintenance Troubleshooting 4 4 19 ajs17f wmf Figure 4 3 Test Point Location Identification...

Страница 66: ...5 220A Instruction Manual 4 20 ajs18f wmf Figure 4 4 Flowchart Summary of Mainframe and Digital Section Troubleshooting Guide...

Страница 67: ...220A to line power and set the POWER switch to ON 9 9 Is the fan running 13 10 10 Are any of the front panel LEDs lit 11 12 11 Loose fan connections or defective fan Repair as required 8 12 Check the...

Страница 68: ...operate modes be selected Note The jumper wire must be installed between the enabled output terminals 21 Is the OVERCURRENT or OVERCOMPLIANCE LED lit 23 22 22 Defective switch LED or standby operate...

Страница 69: ...following and repair as required 1 Open feedback loop current shunt to A5 Preamplifier 2 Overcurrent detector on A8 PCB 3 Failure status logic on A10 PCB 4 Defective A5 Preamplifier PCB 36 Is OVERCOM...

Страница 70: ...5 220A Instruction Manual 4 24 ajs19f wmf Figure 4 5 Flowchart Summary of Analog Section Troubleshooting Guide...

Страница 71: ...ifier PCB Assembly With the 5220A in the operate mode the DVM should read 15 V dc 11 11 Does the DVM read 15 V dc 13 14 12 Troubleshoot the Turn on 1 circuit starting with the level shifting transisto...

Страница 72: ...nd R13 for open circuits If these are OK check the Relay 1 drive signal from the A10 Logic PCB Assembly 23 Is the DVM reading greater than 0 010 V dc 24 25 24 The A6 Driver or A7 Output stage is defec...

Страница 73: ...components may be ordered directly from the manufacturer by using the manufacturers part number or from the Fluke Corporation and its authorized representatives by using the part number under the head...

Страница 74: ...rs listed below or visit us on the World Wide Web www fluke com 1 800 443 5853 in U S A and Canada 31 40 267 8200 in Europe 1 425 356 5500 from other countries Note This instrument may contain a Nicke...

Страница 75: ...List of Replaceable Parts Service Centers 5 5 3 A11 MIS Bus Interface PCB Assembly 491266 C A13 Output Termination PCB Assembly 491274 B...

Страница 76: ...486217 1 B1 Venturi Muffin Fan 4 5 in 103374 1 C1 C8 Capacitor AL 50 000 F 100 10 15 V Hi Term 423525 8 C9 Capacitor CER 0 05 F 80 20 1 000 V Z5U 355420 1 CR1 CR3 Diode SI 50 PIV 20 0 A 483206 2 CR2...

Страница 77: ...Screw Machine PH P Steel 6 32 x 0 625 152181 8 H22 Screw Machine PH P Steel 6 32 x 1 500 114181 4 H23 Screw Machine PH P Steel 8 32 x 0 375 114124 14 H24 Screw Machine PH P Steel 10 32 x 0 625 114066...

Страница 78: ...Snap in Nylon 0 250 ID 102996 1 MP30 Terminal Insulated Standoff Bifurcated 271650 2 MP31 Power Plug Panel 6 A 250 V 3 wire 284166 1 MP32 Card Guide Nylon 4 50 x 0 076 2 snap 256461 12 MP33 Binding Po...

Страница 79: ...or CC 100 10 2 W 109934 2 TM1 5220A Instruction Manual 491936 1 VR1 VR22 Zener Transient Suppressor 7 5 V 10 904896 2 W1 Line Cord 5 15 IEC 3 18 AWG SVT 7 5 ft 284174 1 W2 Cable Fan Power 491308 1 W3...

Страница 80: ...5 220A Instruction Manual 5 8 ajs20f wmf Figure 5 1 5220A Final Assembly...

Страница 81: ...List of Replaceable Parts Service Centers 5 5 9 ajs21f wmf Figure 5 1 5220A Final Assembly cont...

Страница 82: ...5 220A Instruction Manual 5 10 ajs22f wmf Figure 5 1 5220A Final Assembly cont...

Страница 83: ...List of Replaceable Parts Service Centers 5 5 11 ajs23f wmf Figure 5 1 5220A Final Assembly cont...

Страница 84: ...5 220A Instruction Manual 5 12 ajs24f wmf Figure 5 1 5220A Final Assembly cont...

Страница 85: ...List of Replaceable Parts Service Centers 5 5 13 ajs25f wmf Figure 5 1 5220A Final Assembly cont...

Страница 86: ...5 220A Instruction Manual 5 14 ajs26f wmf Figure 5 1 5220A Final Assembly cont...

Страница 87: ...List of Replaceable Parts Service Centers 5 5 15 ajs27f wmf Figure 5 1 5220A Final Assembly cont...

Страница 88: ...5 220A Instruction Manual 5 16 ajs28f wmf Figure 5 1 5220A Final Assembly cont...

Страница 89: ...1 J10 J85 Connector PCB Edge Rec 0 150 CTR 24 Position 295352 2 J80 Connector PCB Edge Rec 0 150 CTR 22 Position 459883 1 MP1 Cover AC Switch 475681 1 MP2 Pushbutton SML Rect C L Repl Green 419747 1...

Страница 90: ...crew Machine PH P Steel 6 32 x 0 250 152140 8 MP1 Transformer Plate Inboard 489468 1 MP2 Transformer Plate Outboard 489476 1 MP3 Forward Transformer Terminator Assembly 660217 1 MP4 Aft Transformer Te...

Страница 91: ...500793 2 H14 Nut Machine Hex BR 28 110619 9 H15 Washer Shoulder Fiber 0 250 ID 110833 4 H16 Screw Machine PH P Steel 10 32 x 0 375 159749 5 H17 Nut Machine Hex Steel 8 32 110544 10 H18 Screw Machine P...

Страница 92: ...5 220A Instruction Manual 5 20 ajs38f wmf Figure 5 4 A3 Capacitor Bus PCB Assembly...

Страница 93: ...918 4 H2 Nut Machine Hex Steel 4 40 110635 4 MP1 Heat Dissipator PCB MTG 1 380 2 000 0 500 386235 4 MP2 Cable Tie 4 0 L 0 100 W 0 75 Dia 172080 6 R4 R6 R9 R15 Resistor CF 100 k 5 0 25 W 348920 4 R7 R1...

Страница 94: ...5 220A Instruction Manual 5 22 ajs39f wmf Figure 5 5 A4 Regulator PCB Assembly...

Страница 95: ...6 32 x 0 812 306804 4 MP2 Shield Preamplifier 486001 1 MP3 Spacer Switch Standoff Polyethylene 0 094 285353 2 Q1 h Transistor SI N JFET Hi Voltage TO 92 393314 1 Q2 h Transistor SI NPN Dual TO 5 47809...

Страница 96: ...1 R24 Resistor CF 33k 5 0 25 W 348888 1 TP1 TP14 Terminal Un insulated Feed through Hole Turret 179283 14 U1 h IC Array 5 Transistors NPN 3 Isolated 2 Differentially Connected 248906 1 U2 h IC OP AMP...

Страница 97: ...List of Replaceable Parts Service Centers 5 5 25 ajs40f wmf Figure 5 6 A5 Preamplifier PCB Assembly...

Страница 98: ...Small Signal 218081 3 Q5 Q9 Q10 h Transistor SI PNP Small Signal 402586 3 Q6 Q7 Q11 h Transistor SI N JFET TO 92 Switch 261578 3 Q8 Q12 h Transistor SI NPN Small Signal 346916 2 R1 R3 Resistor CF 1 k...

Страница 99: ...Fluke Part No Total Quantity R28 R29 Resistor CF 6 8 k 5 0 25 W 368761 2 R30 R31 Resistor CC 10 10 0 5 W 108092 2 TP1 TP4 Terminal Un insulated Feed through Hole Turret 179283 4 VR1 VR2 h Zener Uncomp...

Страница 100: ...882 8 MP9 Insulated Part Transistor Aluminum TO 3 477802 4 MP10 Socket Single PCB For 0 018 0 040 Pin 348201 8 MP11 Heat Dissipator W BEO Washer TO 5 407262 1 MP12 Space Transistor Mount DAP 152207 1...

Страница 101: ...List of Replaceable Parts Service Centers 5 5 29 ajs42f wmf Figure 5 8 A7 Output PCB Assembly...

Страница 102: ...5 220A Instruction Manual 5 30 ajs43f wmf Figure 5 8 A7 Output PCB Assembly cont...

Страница 103: ...839 11 R5 R10 R27 R31 R35 R42 Resistor CF 100 k 5 0 25 W 348920 12 R8 Resistor CF 82 k 5 0 25 W 348912 1 R9 Resistor Variable CER 100 k 10 0 5 W 288308 1 R11 Resistor CF 300 k 5 0 25 W 441535 1 R12 R2...

Страница 104: ...5 220A Instruction Manual 5 32 ajs44f wmf Figure 5 9 A8 Analog Control PCB Assembly...

Страница 105: ...ED Red PCB mount Luminous Intensity 0 5 MCD 369777 9 MP1 Cover Pushbutton Switch 401299 3 MP2 Switch Part SPST Spring 414516 3 MP3 Switch Part SPST Fixed Contact 416875 3 MP4 Pushbutton Square Dark Pu...

Страница 106: ...stor CF 100 k 5 0 25 W 348920 3 R22 Resistor CF 10 k 5 0 25 W 348839 1 R23 R25 Resistor CF 4 7 k 5 0 25 W 348821 2 TP1 TP2 Terminal Un insulated Feed through Hole Turret 179283 2 U1 U4 U5 U20 Isolator...

Страница 107: ...List of Replaceable Parts Service Centers 5 5 35 ajs30f wmf Figure 5 11 A10 Logic PCB Assembly...

Страница 108: ...412700 2 MP2 Connector ACC Micro Ribbon Spring 412718 2 MP3 Spacer Swaged Round Brass 4 40 x 0 187 335604 2 MP4 Spacer Swaged Round Brass 6 32 x 0 570 261875 1 R1 R12 Resistor CF 10 k 5 0 25 W 348839...

Страница 109: ...List of Replaceable Parts Service Centers 5 5 37 ajs31f wmf Figure 5 12 A11 MIS Bus PCB Assembly...

Страница 110: ...achine RH SL Steel 6 32 x 2 750 115022 4 H3 Nut Mini Hex SS 6 32 110569 4 H4 Solder Tin Lead Alloy Preform Ring 516989 32 J1 J4 Banana Plug Panel 101543 4 MP1 Top Support Shunt 489906 1 MP2 Bottom Sup...

Страница 111: ...List of Replaceable Parts Service Centers 5 5 39 ajs32f wmf Figure 5 13 A12 Shunt Assembly...

Страница 112: ...ence Designator Description Fluke Part No Total Quantity C1 Capacitor CER 0 025 F 20 100 V Z5W 168435 1 C2 Capacitor CER 0 1 F 80 20 25 V Z5U 369199 1 R1 Resistor MF 40 2 1 0 125 W 100 PPM 245373 1 R2...

Страница 113: ...No Total Quantity H1 Screw Machine SEMS PH P Steel 6 32 0 250 178533 4 H2 Rivet S Tub Oval Steel 0 118 x 0 218 103606 2 J1 Connector PCB Edge REC 0 150 CTR 40 Position 422550 1 J2 Connector ACC PCB Ed...

Страница 114: ...5 220A Instruction Manual 5 42...

Страница 115: ...in the following accessories subsection 6 2 Accessories 6 3 Rack Mounting Kit M08 205 600 The 5220A can be rack mounted in a standard 19 inch equipment rack using Rack Mounting Kit M08 205 600 Use the...

Страница 116: ...5 220A Instruction Manual 6 2 ajs45f wmf Figure 6 1 Rack Mounting Kit...

Страница 117: ...row of mounting holes 3 Install the cabinet section D of the center section C in the equipment rack 4 Pull the center section of the slide out through the front of the equipment rack until it locks in...

Страница 118: ...5 220A Instruction Manual 6 4 ajs46f wmf Figure 6 2 Rack Slide Kit...

Страница 119: ...7 1 Chapter 7 Schematic Diagrams...

Страница 120: ...5220A Instruction Manual 7 2 5220A 1001 Sheet 1 of 2 ajs47f wmf Figure 7 1 Interconnect Diagram...

Страница 121: ...Schematic Diagrams 7 7 3 ajs48f wm Figure 7 1 Interconnect Diagram cont...

Страница 122: ...5220A Instruction Manual 7 4 ajs49f wmf Figure 7 2 Chassis Wiring...

Страница 123: ...Schematic Diagrams 7 7 5 ajs50f wmf Figure 7 2 Chassis Wiring cont...

Страница 124: ...5220A Instruction Manual 7 6 ajs51f wmf Figure 7 3 Power Supply...

Страница 125: ...Schematic Diagrams 7 7 7 ajs52f wmf Figure 7 3 Power Supply cont...

Страница 126: ...5220A Instruction Manual 7 8 ajs53f wmf Figure 7 4 A5 Preamplifier PCB Assembly...

Страница 127: ...Schematic Diagrams 7 7 9 ajs54f wmf Figure 7 4 A5 Preamplifier PCB Assembly cont...

Страница 128: ...5220A Instruction Manual 7 10 ajs55f wmf Figure 7 5 A6 Driver and A7 Output PCB Assemblies...

Страница 129: ...Schematic Diagrams 7 7 11 ajs56f wmf Figure 7 5 A6 Driver and A7 Output PCB Assemblies cont...

Страница 130: ...5220A Instruction Manual 7 12 ajs57f wmf Figure 7 6 A8 Analog Control PCB Assembly...

Страница 131: ...Schematic Diagrams 7 7 13 ajs58f wmf Figure 7 6 A8 Analog Control PCB Assembly cont...

Страница 132: ...5220A Instruction Manual 7 14 ajs59f wmf Figure 7 7 A9 Front Panel PCB Assembly...

Страница 133: ...Schematic Diagrams 7 7 15 ajs60f wmf Figure 7 7 A9 Front Panel PCB Assembly cont...

Страница 134: ...5220A Instruction Manual 7 16 ajs61f wmf Figure 7 8 A10 Logic PCB Assembly...

Страница 135: ...Schematic Diagrams 7 7 17 ajs62f wmf Figure 7 8 A10 Logic PCB Assembly cont...

Страница 136: ...5220A Instruction Manual 7 18 ajs63f wmf Figure 7 9 A11 MIS Bus Interface PCB Assembly...

Страница 137: ...Schematic Diagrams 7 7 19 ajs64f wmf Figure 7 9 A11 MIS Bus Interface PCB Assembly cont...

Страница 138: ...5220A Instruction Manual 7 20...