The frequency readout should be close to 50 or 60 Hz.
Other readouts can be selected: their functions are indicated in the table
below.
SELECTED
MEASURING RESULT
DC
Readout of the DC part of the input signal
ACrms
Readout of the ACrms part of the input signal
AC+DCrms
Readout of true rms-value of the complete input signal.
Hz
Readout of the frequency.
Pulse+
Readout of the duration of the positive-going part of the input
signal (usually a square wave).
Pulse-
Readout of the duration of the negative-going part of the input
signal (usually a square wave).
Duty+
Readout of the duration of the positive-going part of the input
signal (usually a square wave) as a percentage of the period
time.
Duty-
Readout of the duration of the negative-going part of the input
signal (usually a square wave) as a percentage of the period
time.
Peak max
Readout of the maximum peak value of the input signal.
Peak min
Readout of the minimum peak value of the input signal.
Peak m/m
Readout of peak-to-peak value of input signal.
Crest
Readout of the peak value divided by the rms value.
Phase, Input
[2] only
Readout of the phase difference between the waveforms at input
[1] and input [2]