Clamp
Meters
Safety
Information
5
•
De-energize the circuit or wear personal protective equipment in compliance with
local requirements before you apply or remove the Flexible Current Probe.
•
Measure a known voltage first to make sure that the Product operates correctly.
•
Limit operation to the specified measurement category, voltage, or amperage
ratings.
•
The battery door must be closed and locked before you operate the Product.
•
Connect the common test lead before the live test lead and remove the live test
lead before the common test lead.
•
Remove all probes, test leads, and accessories before the battery door is
opened.
•
Keep fingers behind the finger guards on the probes.
•
Hold the Product behind the tactile barrier.
•
Replace the batteries when the low battery indicator shows to prevent incorrect
measurements.
•
Do not use the HOLD function to measure unknown potentials. When HOLD is
turned on, the display does not change when a different potential is measured.
•
Disconnect power and discharge all high-voltage capacitors before you measure
resistance, continuity, capacitance, or a diode junction.
•
Remove the input signals before you clean the Product.
•
Use only specified replacement parts.