Symbols

 

,

 

Application around and removal from hazardous live 
conductors is permitted. 

T

 

Product is protected by double insulation. 

W

 

Risk of Danger.  Important information. See Instruction Sheet. 

X

 

Hazardous voltage. 

)

 

Conforms to relevant Canadian Standards Association 
directives. 

P

 

Conforms to relevant European Union directives. 

Safety Specifications 

Category Rating: CAT III 1000 V and CAT IV 600 V per 
EN61010-1, Pollution Degree 2      
EMC: EN 61326-1, FCC for emission and immunity  

)

Tested to US and Canadian standards for compliance to 

UL61010-1 and CAN/CSA C22.2 No. 101.1:2004 

P

IEC 61010-1 2

nd

 Edition   IEC 61010-02-032 

Electrical Specifications 

Reference Conditions: 23 

±

 5 

°

C, 20 to 75 % RH; conductor 

centered in jaw opening; no DC component; no adjacent 
conductor. 

Output: 5 mV/A 

Range Frequency 

Accuracy 

±

 (% of r A) 

Crest 

Factor 

2 A to 400 A 

45 Hz to  

400 Hz 

 1.41 

2 A to 300 A 

 4 

300 A to 400 A 

2 + 0.04 

 3 

300 A to 400 A 

45 Hz to  

60 Hz 

9 + 0.04 

> 3 and 

 4 

 

 

Phase Shift 

 

2 A to 5 A 

 4 

°

 

5 A to 40 A 

 3 

°

 

40 A to 400 A 

45 Hz to 

400 Hz 

 2 

°

 

 

 
Typical Bandwidth:
 5 Hz to 20 kHz 
Working Voltage: 1000 V ac rms, in compliance with 
EN61010 
Common Mode Voltage: 1000 V ac rms from earth ground, in 
compliance with EN61010 
Input Load Impedance (of host instrument): > 1 M 

 in 

parallel with up to 47 pF 
Maximum Non-destructive Current: 1000 A 
Duty Cycle: 2 A to 400 A continuous 
Influence of Adjacent Conductor: < 9.0 mA/A 
Influence of Conductor Position in Jaw Opening: 

±

 1.0 % of 

r 0.05 A 

General Specifications 

Output Cable Length: 2.5 m 
Maximum Conductor Size
: 32 mm 
Storage Temperature: -20 

°

C to 60 

°

Operating Temperature: 0 

°

C to 50 

°

Relative Humidity: 10 

°

C to 30 

°

C: 95 % 

 30 

°

C to 40 

°

C: 75 % 

 40 

°

C to 50 

°

C: 45 % 

Temperature Coefficient: 0.01 % X (specified accuracy)/ 

°

C  

(< 18 

°

C or > 28 

°

C) 

Altitude: Operating: 2000 m;  2000 m to 4000 m, derate 
category rating to 1000 V CAT II/600 V CAT III, Non-operating: 
12000 m 
Dimensions: 150 x 70 x 30 mm 
Weight: 114 g 

Measurement Considerations 

Observe the following guidelines when making measurements, 
refer to Figure 1: 

 

Center the conductor inside the Current Clamp jaw. 

 

Make sure the clamp is perpendicular to the conductor. 

 

For optimal reading, make sure the conductor is positioned 
between the alignment marks on the jaws of the Current 
Clamp. 

 

Avoid taking measurements close to other current-carrying 
conductors. 

Operation 

To use the Current Clamp: 
1.

 

Connect the clamp cable to the measurement 
instrument’s current inputs. 

2.

 

Make sure the measurement instrument is set to the 
proper range.  

3.

 

Connect the current clamp jaws around the conductor 
to be measured. 

XW

 Warning 

To avoid shock or personal injury, keep fingers 
behind the tactile barrier, see Figure 1. 

Single Current 
Carrying 
Conductor

Tactile Barrier

Alignment 
Marks

Load Direction Arrow

Connector

 

bbu01.eps 

Figure 1.3140R Setup 

1.800.868.7495

[email protected]

Fluke-Direct

.ca