Symbols
,
Application around and removal from hazardous live
conductors is permitted.
T
Product is protected by double insulation.
W
Risk of Danger. Important information. See Instruction Sheet.
X
Hazardous voltage.
)
Conforms to relevant Canadian Standards Association
directives.
P
Conforms to relevant European Union directives.
Safety Specifications
Category Rating: CAT III 1000 V and CAT IV 600 V per
EN61010-1, Pollution Degree 2
EMC: EN 61326-1, FCC for emission and immunity
)
:
Tested to US and Canadian standards for compliance to
UL61010-1 and CAN/CSA C22.2 No. 101.1:2004
P
:
IEC 61010-1 2
nd
Edition IEC 61010-02-032
Electrical Specifications
Reference Conditions: 23
±
5
°
C, 20 to 75 % RH; conductor
centered in jaw opening; no DC component; no adjacent
conductor.
Output: 5 mV/A
Range Frequency
Accuracy
±
(% of r A)
Crest
Factor
2 A to 400 A
45 Hz to
400 Hz
≤
1.41
2 A to 300 A
≤
4
300 A to 400 A
2 + 0.04
≤
3
300 A to 400 A
45 Hz to
60 Hz
9 + 0.04
> 3 and
≤
4
Phase Shift
2 A to 5 A
≤
4
°
5 A to 40 A
≤
3
°
40 A to 400 A
45 Hz to
400 Hz
≤
2
°
Typical Bandwidth: 5 Hz to 20 kHz
Working Voltage: 1000 V ac rms, in compliance with
EN61010
Common Mode Voltage: 1000 V ac rms from earth ground, in
compliance with EN61010
Input Load Impedance (of host instrument): > 1 M
Ω
in
parallel with up to 47 pF
Maximum Non-destructive Current: 1000 A
Duty Cycle: 2 A to 400 A continuous
Influence of Adjacent Conductor: < 9.0 mA/A
Influence of Conductor Position in Jaw Opening:
±
1.0 % of
r 0.05 A
General Specifications
Output Cable Length: 2.5 m
Maximum Conductor Size: 32 mm
Storage Temperature: -20
°
C to 60
°
C
Operating Temperature: 0
°
C to 50
°
C
Relative Humidity: 10
°
C to 30
°
C: 95 %
30
°
C to 40
°
C: 75 %
40
°
C to 50
°
C: 45 %
Temperature Coefficient: 0.01 % X (specified accuracy)/
°
C
(< 18
°
C or > 28
°
C)
Altitude: Operating: 2000 m; 2000 m to 4000 m, derate
category rating to 1000 V CAT II/600 V CAT III, Non-operating:
12000 m
Dimensions: 150 x 70 x 30 mm
Weight: 114 g
Measurement Considerations
Observe the following guidelines when making measurements,
refer to Figure 1:
•
Center the conductor inside the Current Clamp jaw.
•
Make sure the clamp is perpendicular to the conductor.
•
For optimal reading, make sure the conductor is positioned
between the alignment marks on the jaws of the Current
Clamp.
•
Avoid taking measurements close to other current-carrying
conductors.
Operation
To use the Current Clamp:
1.
Connect the clamp cable to the measurement
instrument’s current inputs.
2.
Make sure the measurement instrument is set to the
proper range.
3.
Connect the current clamp jaws around the conductor
to be measured.
XW
Warning
To avoid shock or personal injury, keep fingers
behind the tactile barrier, see Figure 1.
Single Current
Carrying
Conductor
Tactile Barrier
Alignment
Marks
Load Direction Arrow
Connector
bbu01.eps
Figure 1.3140R Setup