Fluke 3000A Flex 4 Скачать руководство пользователя страница 2

WX

 

Warning 

The Probe may only be used and handled by 

qualified personnel. To avoid electric shock or 

personal injury, follow these precautions: 

 

Do not apply to or remove from hazardous 

live conductors without taking additional 

protective measures.  

 

High voltages and currents may be present in 

adjacent circuits under test. 

 

Do not use the Probe if damaged. Always 

connect to display device before it is installed 

around the conductor. 

 

Use the Probe only as specified in the 

operating instructions; otherwise the safety 

features may not protect you. 

 

Adhere to local and national safety codes. 

Individual protective equipment must be used 

to prevent the shock and arc blast injury 

where hazardous live conductors are 

exposed. 

 

Before each use, inspect the Probe. Look for 

cracks or missing portions of the housing or 

output cable insulation. Also look for loose or 

weakened components.  

 

Use caution when working with voltages 

above 60 V dc, 30 V ac rms or 42 V ac peak. 

Such voltages pose a shock hazard.  

 

Equipment is considered to be used in 600 V 

CAT IV and 1000 V CAT III environment.  
CAT IV equipment is designed to be used at 

the source of distribution system up to the 

energy-meter within the installation. CAT III 

equipment is designed to be used in 

distribution panels, feeders and short branch 

circuits, and the lighting system in large 

buildings. 

 

Do not use Probe in damp or wet 

environments or in locations that hazardous 

gases exist. 

Operating Instructions 

1.  Connect the Probe to the product current input jack.  
2.  Wear protective gloves or de-energize the circuit and 

place the Probe around the conductor under test. Re-
energize the circuit. 

3.  Observe and take measurements as required. 

Positive output indicates that the current flow is in the 

direction shown by the arrow on the Probe. 

4.  Wear protective gloves or de-energize the circuit 

before removal of the Probe. 

Cleaning 

Clean the Probe periodically by wiping it with a damp cloth and 

detergent. Do not use abrasive cleaners or solvents. Do not 

immerse the Probe in liquids. 

Symbols

 

~

 

Do not dispose of this product as unsorted municipal 
waste. Go to Fluke’s website for recycling information. 

-

 

Do not apply around or remove from HAZARDOUS LIVE 
conductors, without using protective gloves. 

T

 

Product is protected by double insulation. 

W

 

Risk of Danger.  Important information. See Instruction 
Sheet. 

X

 

Hazardous voltage. 

)

 

Conforms to relevant Canadian Standards Association 
directives. 

P

 

Conforms to relevant European Union directives. 

Electrical Characteristics 

Input ranges I

I

 L1, L2, L3, N:

 30 / 300 / 3000 / 6000 A AC

 

Measuring range: 

0.8 A to 6000 A AC

 

Intrinsic error

: <2 % of II

 

Position influence: 

max. 

±

2 % of m.v. for distance conductor 

to measuring head >30 mm (1.18 in)

 

Stray field influence: 

<

±

2 A for Iext= 500 A AC and distance 

to measuring head >200 mm (7.87 in)

 

Temperature coefficient:

 <0.05 % / K

 

Current transformer: 

ratio : 

999 kA / 

II

 

Ratio selection: 

by job programming for 174x series; by setup 

menu for 1735

 

Connection: 

3-phase, 3-phase + N; 2-phase L1 and L3  

(2W-meter-method); 7 pole connector 
 

Specifications 

General Specifications 

Cable length:

 4 m (0.15 in) 

Length of measuring head:

 122 cm (48 in) 

Operating temperature range:

 -10 

°

C to +70 

°

C  

(14 

°

F to 158 

°

F) 

Storage temperature range:

 -20 

°

C to +90 

°

C  

(-4 

°

F to 198 

°

F) 

Operating humidity (non-condensing):

 10 % to 80 %

 

Weight:

 1.5 kg (3.3 lb) 

Reference Conditions 

Environment temperature range

: +18 

°

C to +26 

°

C  

(64.4 

°

F to 78.8 

°

F) 

Humidity:

 20 to 75 % rh  

Altitude:

 max. 2000 m (1.24 mi) 

Current:

 nominal value In, sinusoidal waveform, 48 Hz to 

65 Hz, distortion factor: <1 %, no DC component, stray field 
<40 A/m, conductor centered within the Probe. 

Safety Standards 

IEC/EN61010-1:2001 
IEC/EN61010-2-032:2002 
IEC/EN61010-031:2002 

Safety Specifications 

Category Rating: 

1000 V CAT III, 600 V CAT IV, pollution 

degree 2.

 

The application of the Probe on 

uninsulated 

conductors

 is limited to 1000 V ac rms or dc to ground and 

frequencies below 1 kHz.

 

EMC Standards: 

IEC/EN 61326-1:2006

 

 