True-rms Thermal Multimeter
Required Equipment
7
Required Equipment
Table 2 is a list of required equipment for performance tests and calibration adjustment.
Performance Tests
XW
Warning
To prevent possible electrical shock, fire, or personal injury, do not start the
performance test procedures unless the Product is fully assembled.
The performance tests verify the full operation of the Multimeter and measure the accuracy of each function against
product specifications. If the Multimeter fails a part of the test, calibration adjustment or repair is necessary. See
Display Test
To verify that the display operates correctly:
1.
Turn on the Multimeter.
2.
Verify that all the icons show correctly on the display.
If not, repair is necessary. See
Backlight Test
To verify that the backlight operates correctly:
1.
Turn on the Multimeter.
2.
Push
.
The backlight becomes brighter. If not, repair is necessary. See
Table 2. Required Equipment
Recommended Equipment
Measurement Function
Accuracy
5522A Multi-Product Calibrator
(or equivalent)
DC Volts
10 mV to 600 V
±
0.02 %
AC Volts
6 mV to 600 V
±
0.25 % @ 45 Hz to 1 kHz
Resistance
0
Ω
to 5 M
Ω
±
0.225 %
10
Ω
to 30 M
Ω
±
0.375 %
Capacitance
9
µ
F to 900
µ
F
±
0.475 %
Frequency
2 V @ 50 kHz
±
0.025 %
4180 Precision IR Calibrator
-5 °C to +120 °C
±
1 °C
4181 Precision IR Calibrator
35 °C to +190 °C
±
1 °C
Test Lead (PN 650892)
2 mm to 4 mm
NA
Test Lead (PN 1903307)
NA
NA