Input and Channel Configuration
Channel Configuration
3
3-21
Thermistor Channels
Refer to Table 3-8 to configure a thermistor channel.
Table 3-8. Thermistor Channel Setup
1
2
3
4
5
6
hce022.eps
Item Function
Thermistor function selection.
Select the thermistor type. Types available: 2.252 k
Ω
, 5 k
Ω
, and 10 k
Ω
.
Set up the channel for a 2-wire measurement connection. See “Sense Input Configuration” on
page 3-3.
Input a custom alpha-numeric label to help identify the channel (optional).
Set an alarm for this channel (optional). See “HI and LO Channel Alarms” on page 3-30.
Apply Mx+B scaling to the measurement (optional). See “Mx+B Scaling” on page 3-29.
1.888.475.5235