Model 187 & 189
Users Manual
7-10
Function
Range
Resolution
Accuracy
Frequency
500.00 Hz
0.01 Hz
1
±
(0.0050 % + 1)
5.0000 kHz
0.0001 kHz
50.000 kHz
0.001 kHz
999.99 kHz
0.01 kHz
Duty Cycle
10.00% to 90.00 %
0.01 %
±
((voltage range/input voltage) X 300 counts)
5,6
Pulse Width
499.99 ms
0.01 ms
±
(3 % X (voltage range/input voltage) + 1 count)
5,6
999.9 ms
0.1 ms
Temperature
−
200 to +1350
°
C
0.1
°
C
±
( 1% of r 1
°
C)
2,3
−
328 to +2462
°
F
0.1
°
F
±
( 1% of r 1.8
°
F)
2,3
MIN MAX AVG
Response: 100 ms to 80 %
Specified accuracy
±
12 counts for changes > 200 ms
in duration. (
±
40 counts in AC for changes > 350 ms
and inputs > 25 % of range)
FAST MN MX
250
µ
s
4
Specified accuracy ±100 counts up to 5,000 count (full
range) reading. For higher peak readings (to 20,000
counts), specified accuracy
±
2% of reading.
1. Reading will be 0.00 for signals below 0.5 Hz.
2. Accuracy specification is relative to the user-adjustable temperature offset, and assumes ambient temperature stable to
±
1
°
C.
3. For ambient temperature changes of
±
5
°
C, rated accuracy applies after 1 hour.
4. For repetitive peaks; 2.5 ms for single events. Use DC function settings below 20 Hz. 50 mV range not specified.
5. Frequency greater than 5 Hz, except for VDC, 500 mVDC and 3000 mVDC functions; 0.5 Hz to 1 kHz. Signals centered around
trigger levels.
6. Range/input ratios also apply to current functions. 500 counts or 5 % for 10 A ranges.