1652C/1653B/1654B Calibration
Manual Supplement
4/17 1
Change #1
On page 20, replace step 4 with:
4. Place PN 1884378 on the touch pad, then touch with the probe as shown in Figure 4.
On page 21, replace Figure 4 with:
280V
RMS
MAX
280V
RMS
MAX
CAT I
1000V
CAT II
600V
RMS MAX
20V PK
30A
RMS
MAX 20V PK
LO
LO - SENSE
, HI , mA
V
1500V PK
MAX
50V PK
MAX
20V PK
20V PK
N
L
PE
L1
L2
L3
H
E
S
OUTPUT
ZL, Z GND, RCD
HI
LO
HI
LO
METER
METER
5320A MULTIFUNCTION ELECTRICAL
V
A
COM
INPUT
Fluke 5320A
Fluke 1653
PE
(L2/Green)
L
(L1/Red)
PN 1884378
gad002.eps
Figure 4. Touch Pad Sense Test