The AVERAGE parameter determines whether averaging is disabled (OFF) or
enabled (ON). Use the
LR
buttons to select ON or OFF for AVERAGE and
press
ENTER
.
The COUNT parameter determines the number of raw measurements that are
averaged to produce the displayed measurement. The range is from 1 to 10. Use
the numeric buttons to enter a value for COUNT and press
ENTER
.
SET THE INPUT AVERAGING
AVERAGE: OFF
COUNT: 1
5.2
Probe Menu
The
PROBE
menu provides functions for selecting the type of temperature
conversion and setting the characterization coefficients for input channels. The
soft-key functions that appear in this menu are
EDIT PROBE
,
COPY
PROBE
, and
TEST CONV
. If the PROBE password option is set ON, you
must enter the correct password in order to access these functions (see Section
5.5.3). Use the numeric keys and
ENTER
to enter the four-digit password. If
the correct password is entered the
PROBE
menu appears.
PASSWORD REQUIRED FOR ACCESS.
PASSWORD: 0
5.2.1
Edit Probe
The
EDIT PROBE
soft-key allows you to enter probe characterization coeffi-
cients and other probe parameters for temperature calculation. First, you must
select the input channel. A window appears showing a list of input channels.
Each channel in the list is identified with the device name, the channel number,
33
5 Soft-Key Functions
Probe Menu
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...