8.4
Calibration
This section explains the calibration procedure for the 2561 HTPRT Module.
8.4.1
Calibration Parameters
The resistance measurement is derived from the ratio between the voltage mea-
surements of the input resistor and the internal reference resistor (10
Ω
) using
the following equation.
resistance
ADJ ratio
ADJ
ratio
=
+
+
−
( .
_
)
_
( .
)
10 0
1 0
10
0
Two adjustable calibration parameters are used for calibration:
0_ADJ
and
10_ADJ. 0_ADJ
directly affects the measurement at 0
Ω
. It has negligible effect
at 10
Ω
.
10_ADJ
directly affects the measurement at 10
Ω
. It has negligible ef-
fect at 0
Ω
. Each of the parameters has a positive effect at their primary
resistances: increasing the value of the parameter increases the measured resis-
tance. The default and theoretically normal value for each is 0.
8.4.2
Front-panel Access
The calibration parameters
0_ADJ
, and
10_ADJ
can be adjusted to optimize the
accuracy. They can be accessed from the front panel of the 1560 using the
CAL
DEVICE
soft-key function. This is found in the
MODULE
sub-menu.
Pressing the
CAL DEVICE
soft-key shows a list of devices with the module
position number. Use the
UD
buttons to move the cursor to the HTPRT mod-
ule and press
ENTER
.
SELECT A DEVICE TO CALIBRATE
>HPRT 1
TCS
2
PRTS 3
After the device is selected a new window appears showing the parameters and
functions available from the device. New values can be entered for the parame-
ters using the numeric buttons and pressing
ENTER
. The
UD
buttons can be
used to move between parameters.
1560 Thermometer Readout
User’s Guide
122
Содержание 1560 Black Stack
Страница 1: ...1560 Black Stack Thermometer Readout User s Guide Rev 932001 Hart Scientific...
Страница 20: ...1560 Thermometer Readout User s Guide 8 Figure 2 System Diagram...
Страница 27: ...15 2 Introduction Data Processing Figure 5 Data Flow...
Страница 70: ...RESET SYSTEM PARAMETERS PRESS ENTER TO CONTINUE PRESS EXIT TO CANCEL 61 5 Soft Key Functions System Menu...