Preparation

This manual covers the P5514B-70M and P5514B-70M-EP. 
The Comparison Test Pump checks pressure-measuring instruments against Master Test 

Gauges.

Note

The terms “Master Test Gauge” and “Gauge” in this document refer to any 

pressure-measuring instrument such as Transfer Standards, Digital Calibrators, and 

Transducers. The Fluke 700G and 2700G Digital Test Gauges are recommended for 

this application.

This system is only as accurate as the Master Test Gauge used. The Master Test Gauge must 

be regularly calibrated to ensure accuracy is maintained.

How to Contact Fluke Calibration

Fluke Corporation operates worldwide. For local contact information, go to our website: 

www.flukecal.com

To register your product, view, print, or download the latest manual or manual supplement, go 

to our website.
Fluke Corporation

P.O. Box 9090

Everett, WA 98206-9090
+1-425-446-5500

[email protected]

.

Safety Information

General Safety Information is in the printed Safety Information document that ships with the 

Product. It can also be found online at 

www.flukecal.com

. More specific safety information is 

listed where applicable.

Содержание P5514B-70M

Страница 1: ...P Hydraulic Comparison Test Pump Users Manual April 2021 English 2021 Fluke Corporation All rights reserved Specifications are subject to change without notice All product names are trademarks of thei...

Страница 2: ...n warranty service contact your nearest Fluke authorized service center to obtain return authorization information then send the product to that service center with a description of the difficulty pos...

Страница 3: ...ion 1 How to Contact Fluke Calibration 1 Safety Information 1 Specifications 2 Operating Fluid Compatibility 2 DUT Connection 3 Operation 4 Priming 4 Pressure Operation 4 Ancillary Equipment 5 Pointer...

Страница 4: ...ii P5514B 70M P5514B 70M EP Users Manual...

Страница 5: ...st Gauge used The Master Test Gauge must be regularly calibrated to ensure accuracy is maintained How to Contact Fluke Calibration Fluke Corporation operates worldwide For local contact information go...

Страница 6: ...ion To avoid damage to the instrument check the quality of the operating fluid during use If the fluid becomes discolored cloudy or particles appear in the reservoir drain and flush the system with cl...

Страница 7: ...apter Check that the sealing face of the device to be fitted is clean and undamaged as scratches or dents can form leak paths To connect the DUT 1 Select the appropriate adapter from the adapter set t...

Страница 8: ...during priming operation Pressure Operation Follow these steps 1 Turn the screw press in to generate pressure 2 Allow a few moments for the system to stabilize before taking any readings especially a...

Страница 9: ...will shock the system which can damage the sensitive instruments attached to the system Ancillary Equipment Pointer Remover Punch P5551 To remove and refit the pointer of a pressure gauge use the poi...

Страница 10: ...ruction manuals for the gauges in Table 2 for detailed lists of available ranges and specifications Table 2 Gauges Figure 2 Fluke 700G Model Nominal Specification 700G See Figure 2 0 05 FS 700RG 0 04...

Отзывы: