8-1
Chapter 8
Ancillary Equipment
Dirt/Moisture Trap, P5532
If there is any doubt that the instrument to be tested is not internally clean, then the
addition of the P5532 Dirt/Moisture Trap will protect the deadweight tester from
contamination and possible damage. Any particles or moisture present within the DUT
will fall and be trapped in the aluminum chamber, which can be periodically inspected by
the operator.
The top connection reproduces the test port on the deadweight tester, allowing the
standard gauge adapters to be used.
gmg14.bmp
Figure 8-1. Dirt/Moisture Trap