Multi-Product Calibrator
How to Select Line Voltage
7
100V
/
120V
220V
/
240V
T5.0A 250V
(SB)
T2.5A 250V
(SB)
/
63Hz
47Hz
600VA MAX
WARNING:
TO AVOID ELECTRIC SHOCK GROUNDING
CONNECTOR IN POWER CORD MUST BE CONNECTED
FUSE
MAINS SUPPLY
CHASSIS
GROUND
WARNING:
TO AVOID PHYSICAL INJURY, INSURE THA
IS PROPERLY INSTALLED BEFORE ENERGIZING INSTRU
CAUTION
FOR FIRE PROTECTION REPLACE O
WITH A 250V FUSE OF INDICATED RATING
0V
(SB)
120
120
240
Changing Line Fuse
Changing Line
Voltage
Line Voltage
Indicator
gjh004.eps
Figure 3. How to Access the Fuse and Select Line Voltage
How to Select Line Voltage
The calibrator arrives from the factory configured for the line voltage normally
appropriate for the country of purchase, or as specified at the time of your purchase order.
You can operate the 5522A Calibrator from one of four line voltage settings: 100 V,
120 V, 200 V, and 240 V (47 Hz to 63 Hz). To check the line voltage setting, note the
voltage setting visible through the window in the power line fuse compartment cover
(Figure 3). The allowed line voltage variation is 10 % above or below the line voltage
setting.
To change the line voltage setting, complete the following procedure:
1.
Disconnect line power
.
2.
Open the fuse compartment by inserting a screwdriver blade in the tab located at the
left side of the compartment and gently pry until it can be removed with the fingers.
3.
Remove the line voltage selector assembly by gripping the line voltage indicator tab
with pliers and pulling it straight out of its connector.
4.
Rotate the line voltage selector assembly to the desired voltage and reinsert.
5.
Verify the appropriate fuse for the selected line voltage (100 V/120 V, use 5 A/250 V
slow blow; 220 V/240 V, use 2.5 A/250 V slow blow) and reinstall the fuse
compartment by pushing it back into place until the tab locks.