Multifunction Electrical Tester Calibrator
Verification
65
a. Connect the ground bond/loop/line resistance decade output terminals PE and N to the
8588A. Use a four-wire connection and set the 8588A for four-wire measurement.
b. Select the
Z GND
function on the Product. Push the
Mode
softkey and select
OPEN
.
c. Push
O
.
d. The measured resistance reads >100 k
Ω
.
Figure 38. Ground Bond Resistance and Loop/Line Resistance Value Connections
Fluke 5322A
Fluke 8588A
2-wire Mode
Fluke 5322A
Fluke 8588A
4-wire Mode