5322A
Calibration Manual
48
DC Current Ranges 0.3 A, 3 A, 30 A Adjustment
To adjust these dc current ranges:
1. Change the mode on the 5522A to DCI and connect it to the Product output terminals, see
:
Figure 34. DC Current 0.3A, 3 A, 30A Adjustment Connections
2. Set the first current adjustment point to 300 mA low (0 mA) and push the
Select
softkey.
3. Switch the output terminals ON. The Product reads approximately the nominal value.
4. Push the
Write
softkey to save new value.
5. Go through and adjust all points in the table.
6. When the process finishes, push the
Exit
softkey to go to the basic calibration menu.
This completes this part of the adjustment.
1
4
7
/
+
STBY
5522A CALIBRATOR
Fluke 5322A Fluke 5522A