Multifunction Electrical Tester Calibrator
Adjustment Procedures
41
Ammeter Adjustment HVR I01
This procedure requires you to set one constant at 20 mA dc to adjust the test voltage auxiliary
voltmeter.
1. Connect the Product to the OUTPUT HI
Ω
HI and LO terminals, see
.
Figure 30. High Resistance Meters 1.5 kV Adjustment Connections 2
2. Select
HVR I01 Test current 20mAdc (Short)
from the calibration menu and push the
Select
softkey.
3. Set the output current on the 5522A to 20 mA dc and switch the output terminals ON.
4. Use the rotary knob or cursor keys to set the value to 20.0 mA.
5. Push the
Write
softkey.
This completes this part of the adjustment.