5322A
Calibration Manual
34
Low Resistance Source Meters Adjustment
To adjust the low resistance meters, see
:
1. Connect the 5522A to the OUTPUT HI
Ω
HI and LO terminals.
Figure 25. Low Resistance Source Meters Adjustment Connections 1
2. Select
Low Resistance Meters
from the calibration menu and push the
Select
softkey.
3. Select
LVR I01 Current 1Adc
and push the
Select
softkey.
4. Set the output current on the 5522A to 1 A dc and switch the output terminals ON.
5. Use the rotary knob or cursor keys to set the value of the Product to show 1.00 A.
6. Push the
Write
softkey to confirm the new adjustment value.
1
4
7
/
+
STBY
5522A CALIBRATOR
Fluke 5322A Fluke 5522A