background image

Theory and Applications

Applications 

3

 

 

3-5

 

Table 3-4.  kVp vs. Time Study at 300 mA 

mA 

T (sec) 

ISF X FDD

MAX

FDD (Inches) 

Measured kVp 

300 0.500  1.000 

36.0 

 

300 0.250  0.707 

25.5 

 

300 0.125  0.500 

18.0 

 

300 0.067  0.365 

13.1 

 

 

mAs Reciprocity Corrections 

A mAs reciprocity examination of an x-ray system involves variation of the mA or time or both 
simultaneously.  An ionization chamber (RAD-CHECK Plus, Model 06-526) is used to measure the output 
of the machine and mR values are then calculated.  The mR values should be proportional to the mAs 
quantity. Normally the kV is held constant on the generator. 

As discussed above, the kV may not be a constant function of mA and time if there is a malfunction in the 
machine. 

When mAs reciprocity is being tested, noted variations in the mR value may be due to kV fluctuations as 
well as mA and timer nonlinearities. A simple correction for mAs reciprocity fluctuations due to kV 
changes can be made if the kV fluctuations are known. If the percent change in kV is small, the mR 
output will change as the square function of the kVp. That is, the percent change in mR output is double 
the percent change in kV over a small percentage range (Table 3-5). 

Table 3-5.  mAs Reciprocity 

Measured kVp 

mA 

Time 

mAs

 

mR mR/mAs 

80 100 

0.1 

s  10 

160  16.0 

82 300 

0.1 

s  30 

525  17.5 

 
In Table 3-5 there is a 9.4% increase in mR/mAs values. There is a 2.5% increase in kVp value which 
would contribute to an approximate 5% increase in mR output. The 94% non-linearity in mAs reduces to 
9.4% -5% = 4.4% which is acceptable. 

From this analysis, it is evident that the mA/kV compensation needs correction, whereas, without this 
analysis, the mA or time looks suspect. If the changes in kV and mAs are in the opposite direction, the 
mAs reciprocity results from the mR values may be very nearly linear and a simple reciprocity check 
would indicate that the machine seems to be functioning properly. However, the use of the kVp Meter 
would indicate the kV error and possibly an error in the mAs which was being cancelled out by the kV 
error. 

Waveform Diagnostics 

If the mA is held strictly constant, and filtration does not change through the exposure, then variations in 
the radiation waveform can be interpreted as variations in the kV waveform. It is difficult to reduce the kV 
waveform from a single detector waveform even if the mA is held strictly constant. Relative judgments can 
be made from the radiation waveform; however, absolute judgments should be avoided.  For example, 
the maximum observed point on a radiation waveform corresponds to the maximum kV point in the 
waveform, but it may also be closely aligned to a maximum in the mA waveform. 

The radiation waveform can exhibit symptoms of one or more type of malfunctions in an x-ray system as 
listed in to Table 3-6. 

Table 3-6.  Waveform Diagnostics 

Содержание Victoreen 07-494

Страница 1: ...07 494 Wide Range Digital kVp Meter Operators Manual December 2006 Manual No 168001 Rev 4 2006 Fluke Corporation All rights reserved Printed in U S A All product names are trademarks of their respective companies ...

Страница 2: ...Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 www flukebiomedical com rms ...

Страница 3: ...ents 3 2 3 6 Detector Positioning 3 3 3 7 Low Battery 3 3 3 8 Applications 3 3 Section 4 Operation 4 1 4 1 Making a Measurement 4 1 4 2 Interpreting Front Panel Warning Indications 4 2 4 3 Viewing the Waveform Output 4 3 Section 5 Calibration and Adjustments 5 1 5 1 Low Battery Threshold Adjustment 5 1 5 2 End of Shot Adjustment 5 3 5 3 50 90 kVp Calibration 5 3 5 4 80 150 kVp Calibration 5 3 Sect...

Страница 4: ... Blank page ...

Страница 5: ...t which non invasively measures the effective peak potential applied to a tungsten target diagnostic x ray tube It uses two differentially filtered x ray detectors whose ratio of integrated outputs is calibrated over the ranges of 50 to 90 kVp and 80 to 150 kVp The desired range is selected using the range selector on the top of the unit Figure 1 1 Model 07 494 Wide Range Digital kVp Meter ...

Страница 6: ... BAT indicator Below Range kVp Flashing middle decimal point Above Range kVp Flashing left decimal point Above Range mAs Flashing right decimal point Below Range mAs Flashing right decimal point and 00 0 display LED Indicators Auto Reset New exposure detected last kVp reading cleared Range Fault Range select sliding filter not properly engaged Controls Power Switch ON OFF Phase Selection Switch Si...

Страница 7: ...Victoreen 07 494 Operators Manual 1 2 mA kVp Figure 1 2 mAs vs kVp Minimum Requirements Typical Single Phase X ray Unit ...

Страница 8: ...Victoreen 07 494 Operators Manual Blank page ...

Страница 9: ...nstruction Manual d Part No 010023 Registration Card NOTE If any of the listed items are missing or damaged notify Fluke Biomedical at 440 248 9300 2 2 Storage If necessary to store the kVp meter prior to use pack it in the original container if possible Be sure the storage area is free of corrosive materials vibrations and fluctuations in temperature and humidity Also be sure the environmental sp...

Страница 10: ...4 Operation Auto Reset LED The auto reset LED is lit whenever the kVp Meter detects an x radiation The LED will be on for 0 5 seconds or for the duration of the exposure whichever is greater The previous reading is cleared when the LED turns on Range Fault LED The range fault LED is lit whenever the range selector is between ranges or not properly engaged Power Switch ON OFF The power switch is an...

Страница 11: ...ned OFF before removing the batteries BNC Connector A BNC connector located on the rear panel provides a signal from the radiation detection diodes The signal can be observed on an oscilloscope during an exposure However the signal lasts only as long as the exposure Therefore a storage scope or camera is required to view the signal for extended periods of time CAUTION Turn off power switch before ...

Страница 12: ...Victoreen 07 494 Operators Manual Blank page ...

Страница 13: ...differences with respect to the calibration beam With lower filtration the x ray beam is as hard as the calibration beam and the results are lower With more filtration the beam is harder than the calibration beam and the results are higher Refer to Table 3 1 Table 3 1 Filtration Effects Selected Range kVp Display Increase mm AI 50 90 kVp 65 80 0 34 0 40 80 150 kVp 100 120 0 40 0 45 The results in ...

Страница 14: ... 90 18 in 45 cm 36 in 30 cm 80 150 kVp 80 to 100 100 to 150 18 in 45 cm 36 in 30 cm 3 5 mAs Requirements The curves in Figure 1 2 see Section 1 2 Specifications are provided as guidelines for selecting the mAs required for a particular kVp measurement Any mAs beyond the required amount is not used in the measurement To determine the mAs required 1 Locate the kVp to be measured on the horizontal ax...

Страница 15: ...acy Determinations With the Digital kVp Meter it is possible to enter a diagnostic x ray room and within one minute determine kVp accuracy for several stations of the x ray machine As a result a significant cause of poor image quality can be immediately diagnosed without affecting patient throughput Calculating Generator Loading Effects In an x ray generator when the load mA changes the kV may als...

Страница 16: ...at occurs during the tail end of the exposure A greater portion of the total exposure can be included by increasing the mAs requirement by increasing FDD There is no limit on maximum exposure time To implement this application 1 Select the kV the mA and the longest exposure time to be studied The product of the mA and time should correspond to the minimum mAs requirements determined from Figure 1 ...

Страница 17: ...le 3 5 there is a 9 4 increase in mR mAs values There is a 2 5 increase in kVp value which would contribute to an approximate 5 increase in mR output The 94 non linearity in mAs reduces to 9 4 5 4 4 which is acceptable From this analysis it is evident that the mA kV compensation needs correction whereas without this analysis the mA or time looks suspect If the changes in kV and mAs are in the oppo...

Страница 18: ...microseconds Overshoot Problems Waveform exhibits overshoot on the first 2 or 3 pulses 3 phase machine Contactor Bounce Waveform peaks contain noise or smaller peaks riding on major peaks Rectifier Imbalance Relative peak heights are different and repeatable i e one high and one low or in a three phase machine high medium low high medium low high medium low Load Effects Waveform exhibits drooping ...

Страница 19: ...the fluoroscopic mode the loading on the generator is much lighter than normal and the radiation spectrum more closely resembles that produced by a three phase rather than single phase generator 3 Set the range selector switch located on the top panel to the appropriate range for the kV to be measured 4 Set the desired Focus to Detector Distance FDD using the guidelines listed in Table 4 1 Table 4...

Страница 20: ...l indicates that the measured kVp is below the calibrated range that is that the kVp is below 80 kVp with the range selector in the 80 to 150 kVp setting or below 50 kVp with the range selector in the 50 to 90 kVp setting Indication A Explanation Measured kVp calibrated range B C D Display 00 00 Measured kVp calibrated range Beam intensity too high measurement time too short C D Display 00 0 Beam ...

Страница 21: ...detail in Section 3 Theory and Applications A flashing decimal point the C D on the front panel with a displayed value equal to 00 0 indicates that the minimum mAs requirements have not been met Increase either the time or the mA setting or decrease distance to obtain the required results 4 3 Viewing the Waveform Output The BNC connector on the rear panel provides a signal from the unfiltered radi...

Страница 22: ...Victoreen 07 494 Operators Manual Blank page ...

Страница 23: ...lus TP3 4 Connect a voltmeter between battery plus TP3 and analog ground TP4 5 Turn on the power supply and adjust it for 9 V 6 Turn on the kVp Meter The voltmeter should read between 2 4 V and 3 2 V 7 Slowly decrease the power supply voltage until the voltage at the test point starts to drop between 6 5 V and 7 5 V 8 Slowly increase the power supply until the voltage across the test points just r...

Страница 24: ...on 5 2 Adjustments 80 150 kVp Calibration Adjustments R6 d R3 Low Battery Threshold Adjustment R25 Three Phase Adjustments R19 and R60 R TP1 R60 TP1 R R TP3 TP2 Below Batter TP8 R33 TP7 TP4 TP1 TP9 TP5 R19 R Figure 5 1 Adjustment and Test Point Locations ...

Страница 25: ...gle phase or three phase depending on the type of x ray machine used for calibration 2 Set the range selector switch to the 50 90 kVp range 3 Remove the cover from the meter 4 Locate the meter at an FDD of 24 inches 5 Set the x ray machine to 50 kVp 300 mA and 1 5 seconds 6 Make an exposure 7 Adjust R5 until the display reads 50 kVp 8 Set the x ray machine to 85 kVp 300 mA and 0 3 seconds 9 Make a...

Страница 26: ...0 kVp 50 mA and 0 3 seconds 9 Make an exposure NOTE If the C D decimal point is flashing indicating that the intensity of the exposure is too high decrease the mA or increase the FDD and repeat the exposure 10 Adjust R3 until the display reads 140 kVp 11 Repeat Steps 5 through 10 until the readings are accurate ...

Страница 27: ...ollowing procedure to gain access to the circuit board components and test points NOTE Refer to Figure 5 1 see Section 5 Calibration and Adjustments for adjustment and test point locations 1 Remove the two screws from the case bottom 2 Carefully lift the top cover to expose the circuit board 3 If necessary remove the four screws which attach the slide filter to the circuit board to gain access to ...

Страница 28: ...time NOTE Data with a kVp test cassette may be useful Be sure to note positioning error and film usage when using the cassette 4 Record all parameters of the measurement including geometry cable connections tube ID generator ID divider and kVp Meter waveform oscilloscope pictures oscilloscope and voltmeter ID 5 Periodically every 6 months or when a serious discrepancy occurs repeat Steps 1 through...

Страница 29: ...ges Equipment recommended for troubleshooting the kVp Meter includes 1 A digital voltmeter Fluke Model 77 or equivalent 2 A storage oscilloscope Tektronix Model 2430 or equivalent Use the following procedure 1 Connect the DVM negative lead to TP4 circuit ground NOTE NOTE All measurements will be made with reference to TP4 2 Apply power to the unit under test 3 Test for the voltages listed in Table...

Страница 30: ...Victoreen 07 494 Operators Manual 6 4 CH1 TP1 CH2 TP8 Setup Single phase x ray machine 74 5 kVp 300 mA 28 SDD CH1 1V CH2 1V A 20ms 328mV CH1 CH1gnd CH2gnd CH1 CH2 Figure 6 2 Oscilloscope Waveform A ...

Страница 31: ...ce and Troubleshooting Verifying Supply Voltages 6 5 CH1 TP1 CH2 TP9 Setup Single phase x ray machine 74 2 kVp 300 mA 28 SDD CH1 1V A 20ms 563mV CH1 CH2 1V CH1gnd CH2gnd CH1 CH2 Figure 6 3 Oscilloscope Waveform B ...

Страница 32: ...Victoreen 07 494 Operators Manual 6 6 CH1 TP1 CH2 TP2 Setup Single phase x ray machine 75 5 kVp 300 mA 28 SDD CH1 1V A 20ms 563mV CH2 CH2 1V CH1gnd CH2gnd CH1 CH2 Figure 6 4 Oscilloscope Waveform C ...

Страница 33: ... Blank Page ...

Страница 34: ...Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland Ohio 44139 440 498 2564 www flukebiomedical com rms ...

Отзывы: