ESA612
Users Manual
58
Measuring Current
The Analyzer can make dc only, ac only, and ac+dc
current measurements up to 10 mA. To make a current
measurement:
1. Press the softkey labeled
Leakage
from the
Point-To-Point menu.
2. Using
or
select between ac only, dc only, or
ac+dc measurement mode.
3. Insert test leads in the RED (V/
Ω
/A ) and BLACK
jacks.
4. Place the leads on the two points the unknown
current may flow and read the measurement in the
Analyzer’s display.
Simulating ECG Waveforms
The Analyzer is capable of generating various waveforms
on the applied parts connection posts. These signals are
used to test the performance characteristics of ECG
monitors and ECG strip printers. See Figure 34 for proper
connections between the Analyzer and an ECG monitor.
For monitors using the snap style connectors, insert the
BJ2ECG adapter into the connectors at the top of the
Analyzer and connect the monitor leads to the snap
connectors on the adapter.
Note
If the ECG monitor/interpreter has banana posts
instead of snaps, use the optional universal snap
to banana adapter to connect to the Analyzer.
To access the ECG Simulation Waveform menu shown in
Figure 33, press
. From this menu, a number of
different waveforms are selected through F1, and the rate
or frequency of the waveform is selected through F2.
fis109.jpg
Figure 33. ECG Waveform Simulation Menu
Содержание ESA612
Страница 6: ......
Страница 10: ...ESA612 Users Manual iv ...
Страница 12: ...ESA612 Users Manual vi ...
Страница 26: ...ESA612 Users Manual 12 fis113 png Figure 5 DUT Connections to the Analyzer ...
Страница 34: ...ESA612 Users Manual 20 fis112 png Figure 10 Ground Wire Protective Earth Resistance Measurement Connections ...
Страница 36: ...ESA612 Users Manual 22 faw26 png Figure 11 Ground Wire Protective Earth Resistance Measurement Schematic ...
Страница 38: ...ESA612 Users Manual 24 faw17 png Figure 13 Mains to Protective Earth Insulation Resistance Test Schematic ...
Страница 40: ...ESA612 Users Manual 26 faw19 png Figure 15 Mains to Applied Parts Insulation Test Schematic ...
Страница 42: ...ESA612 Users Manual 28 faw21 png Figure 17 Applied Parts to Non Earth Conductive Points Schematic ...
Страница 50: ...ESA612 Users Manual 36 gtv29 png Figure 21 Lead to Ground Patient Leakage Current Test Schematic ...
Страница 52: ...ESA612 Users Manual 38 gtv30 png Figure 23 Lead to Lead Patient Auxiliary Leakage Current Test Schematic ...
Страница 58: ...ESA612 Users Manual 44 faw22 png Figure 25 Alternative Equipment Leakage Current Test Schematic ...
Страница 64: ...ESA612 Users Manual 50 gtv25 png Figure 28 Direct Applied Parts Leakage Current Test Schematic ...
Страница 66: ...ESA612 Users Manual 52 gtv32 png Figure 29 Differential Leakage Current Test Schematic ...
Страница 68: ...ESA612 Users Manual 54 fis120 png Figure 30 1 to 10 Adapter Connections ...
Страница 70: ...ESA612 Users Manual 56 fis121 png Figure 31 ECG Lead Connection with 1 to 10 Adapter ...
Страница 74: ...ESA612 Users Manual 60 fis115 png Figure 34 ECG Monitor Connections ...
Страница 88: ......