ProSim™ 8
Getting Started Manual
18
ECG Artifact
Type
........................................................................ 50 Hz, 60 Hz, muscular, baseline wander, respiration
Size
......................................................................... 25 %, 50 %, 100 % of the normal sinus R-Wave for each lead
Lead Select
............................................................ All, RA, LL, LA, V1, V2, V3, V4, V5, V6
Fetal / Maternal-ECG
Fetal Heart Rate (Fixed)
........................................ 60 BPM to 240 BPM in 1 BPM steps
Fetal Heart Rate (IUP)
........................................... 140 BPM at beginning, then varies with pressure
Intrauterine-Pressure Waveforms
....................... Early deceleration, late deceleration, and acceleration
Wave Duration
....................................................... 90 seconds, bell-shaped pressure curve, from 0 mmHg to 90 mmHg and returning to
0 mmHg
IUP Period
.............................................................. 2 minutes, 3 minutes, or 5 minutes; and manual
Default Settings
..................................................... FHR 140 BPM, early deceleration wave, manual
Invasive Blood Pressure
Channels
................................................................ 2, each independently settable with identical parameters and are individually electrically
isolated from all other signals
Input/output Impedance
....................................... 300
Ω
±
10 %
Exciter Input Range
.............................................. 2.0 V to 16.0 V peak
Exciter-Input Frequency Range
........................... DC to 5000 Hz
Transducer Sensitivity
......................................... 5
μ
V/V/mmHg (default) or 40
μ
V/V/mmHg
Pressure Accuracy
...............................................
±
(1 % of s 1 mmHg) Accuracy guaranteed for DC excitation only
Static Pressure
...................................................... -10 mmHg to +300 mmHg in 1 mmHg steps
Pressure Units
....................................................... mmHg or Kpa