4. MAINTENANCE
4. MAINTENANCE
4.3 Maintenance of oil
4.3.1 Checking the oil-level
During operation of the pump, the oil level must always remain between marks H and L on the oil-
level glass. The amount of oil must be checkedandtoppedup as required
.
Note
: The pump must be switched off before topping up the oil. It is recommended to power off
other components connectedtopump for safety reason.
If gases of liquids dissolved in the oil result in a deterioration of the ultimate pressure, the oil can be
degassed by allowing the pump to run for approx. 30 min. with the intake port closed and the gas
ballast valve open
.
Replace the oil
2,500 operation hours
Inspect and clean the inlet-filter
Every oil change
Clean or replace the gas-ballast o-ring
Every oil change
Clean the motor fan-cover
Yearly
Clean and overhaul the pump
15,000 operation hours
Fit new blades
30,000 operation hours
Test the motor condition
15,000 operation hours
Oil change Frequency
Application
Frequency (hrs)
High Vacuum in Lab environment, or system seldom exposed to the air
2,500
High Vacuum in manufacturing environment, often exposed to the air
1,200
Vacuum furnace, large gas-exhaust diffusion system, system with booster pump
600
Vacuum drier, vacuum molding, vacuum packing
200
Vacuum distillation, low vacuum tank
120
Overhaul
is a process of dismantlement the pump, cleaning all parts and reassembling.
It is highly important to improve the performance and elongate lifetime of the pump.
WSA recommends users to do overhaul at least every other year.
Maintenance Frequency
Operation
Frequency
Check the oil-level
Daily
4
.
3
.
2 Oil change
For proper operation of the pump, it is essential that the pump has an adequate supply of the
correct and clean oil at all times
.
Clean vacuum oil is basically odorless and transparent.
(MR-
200 has pale yellow color.)
As operating pumps, the color of vacuum oil turns from yellow through orange into red brown.
This discoloration is caused by gases and foreign materials pumped in and this mixture would
impairs the vacuum level and damage the interior ofpump.
Change vacuum oil when discoloration is dark yellow to red brown. (4 in above picture.)
When the color reach 6~7 with deterioration, the interior of pump would be damaged and
overhaul process is necessary beforeoperatingthe pump again.
If the vacuum oil becomes cloudy as milk color, it is caused by moisture flowedin.
In this case, stop the pump and wait until the fluid is divided into two layers; vacuum oil and water.
Then drain the water layer completely through oil drainplug.
Design the process so as moisture not to be pumped in and install a cold trap if necessary.
Before and after a prolonged storage, additional change of oil is required.
If the speed of deterioration of oil is too fast, apply dust filter or oil filter to the pump.
(refer to sec 7.4 : WMT-250/400)
Contact WSA for further information.
Caution
Only change the oil after the pump has been switched off and
while the pump is still warm..
Procedure
①
Remove the oil-drain plug and let the used oil drain into a suitable container.
②
When the flow of oil slows down, screw the oil-drain plug back in, briefly switch on the pump
(max. 10 sec) and then switch it off again. Remove the oil-drain plug once more and drain out the
remaining oil.
③
Screw the oil-drain plug back in (check the gasket and reinstall a new one if necessary.)
④
Remove the oil-fill plug and fill in with fresh oil.
⑤
Screw the oil-fill plug back in.
* Recommended Vacuum oil :
NEOVAC MR 200
Caution
We can only guarantee that the pump operates as specified by the
technical data if the vacuum oil recommended by us is used.
13