LACO USER MANUAL - GASCHECK G SERIES
17
© 2018 FLUIDX Equipment, Inc.
3. Ensure the instrument is in clean air and then press the ENTER key. The
instrument will Zero. This will take a few seconds, but once complete a
READY message will appear.
4. Insert the instrument’s probe fully into the sample bag containing
the gas and press the ENTER key. The instrument will automatically
sequence through the cold and hot stages of calibration but when
finished gives a summary of values.
5. Press the ENTER key to return back to the calibration option screen.
Press the ESC key to start using the instrument or press the ENTER key
to calibrate again.
NOTE
To avoid errors in calibration, the following points should
be noted:
• Avoid bending the inner probe as this will affect the GasCheck’s accuracy
• The GasCheck can detect changes in humidity and carbon dioxide. Avoid
breathing on the probe.
• Avoid pressurizing the sample bag. This will cause significant errors in
reading.
5.1.3 CALIBRATION PROCEDURE USING CC/SEC UNITS
Please read this entire procedure before proceeding with this calibration
routine.
1. Check your LACO CalMaster™ Leak Standard to ensure the pressure
reading matches the leak rate shown on the calibration stick. The
pressure gauge may be installed on your leak standard or for open
style leak standards the pressure gauge is set on helium regulator.
2. Select the CUSTOM CAL symbol from the instrument’s menu.
3. Ensure the instrument is in clean air and then press the ENTER key. The
instrument will Zero. This will take a few seconds but once complete, a
READY message will appear.
4. Insert the instruments probe into the CalMaster™ Leak Standard’s
outlet and press the enter key. The instrument will automatically
sequence through the cold and hot stages of calibration. When
finished, it gives a summary of values.
5. Pressing the ENTER key will return back to the calibration option
screen. Press the ESC key to start using the instrument or press the
ENTER key to recalibrate again.
NOTE
The GasCheck can detect changes humidity and carbon
dioxide. Avoid breathing on the probe.
Содержание GasCheck G Series