6. Maintenance
6.1. GENERAL
DIRECTIONS
Mounting, electrical installation, start-up and maintenance of the instrument may only
be carried out by trained personnel authorized by the operator of the facility. Personnel
must read and understand this Operating Manual before carrying out its instructions.
The F126-P-EG may only be operated by personnel who are authorized and trained by
the operator of the facility. All instructions in this manual are to be observed.
Ensure that the measuring system is correctly wired up according to the wiring
diagrams. Protection against accidental contact is no longer assured when the housing
cover is removed or the panel cabinet has been opened (danger from electrical shock).
The housing may only be opened by trained personnel.
Take careful notice of the “ Safety rules, instructions and precautionary measures “ in
the front of this manual.
The F126-P-EG does not require special maintenance unless it is used in low-temperature
applications or surroundings with high humidity (above 90% annual mean). It is customers
responsibility to take all precautions to dehumidify the internal atmosphere of the F126-P-EG in such
a way that no condensation will occur, for example by placing dry silica-gel in the casing just before
closing the enclosure.
Furthermore, is required to replace or dry the silica gel from time to time as advised by the silica gel
supplier.
Battery life-time:
It is influenced by several issues as:
Input frequency: the higher the frequency the shorter the battery life-time.
Display update: fast display update has major influence; SETUP 51.
Pulse output and communication.
Low temperatures; the available power will be less due to battery chemistry.
Note:
It is strongly advised to disable unused functions.
Check periodically:
The condition of the casing, cable glands and front panel.
The input/output wiring for reliability and aging symptoms.
The process accuracy. As a result of wear and tear, re-calibration of the flowmeter might be
necessary. Do re-enter any subsequent Span alterations.
The indication for low-battery.
Clean the casing with soapy-water. Do not use any aggressive solvents as these might damage
the coating.
6.2. REPAIR
This product cannot be repaired by the user and must be replaced with an equivalent certified
product. Repairs should only be carried out by the manufacturer or his authorized agent.