Read the display

1. 

During use the ACS900 monitors each inflation 

to make sure it is delivering prophylaxis to 

the patient. The accumulated therapy time is 

displayed in the bottom right corner.

2.  If the pump senses that a garment has been 

removed from the limb, while the pump is in run 

mode, the corresponding limb graphic flashes.

-XX

 

 

Caregivers are alerted that no prophylaxis is 

being delivered.

The sequence of notifications:

• 

The leg icon starts to flash and the non-

compliance time appears on the screen.

•  After 3 cycles, the warning triangle appears 

on the screen and the green LED starts to 

flash.

• 

After 10 minutes, the flashing green LED turns 

to solid yellow with an optional audible alarm.

Additionally, the pump will count and display 

how long therapy has not taken place (non-

compliance time). This is initially displayed in 

flashing minutes.

-XX

3.  After 1 hour of monitored non-compliance the 

display switches over to show the time in flashing 

hours.

1

4.  Press the right control button to see the additional 

minutes to the hour, a “–” is shown before 

minutes on the display.

-45

Rectify warning/alert

1.  In order to clear the Compliance Monitoring  

warning/alert, refit any removed garments to the 

limb.

2.  The non-compliance time is automatically cleared 

when the pump senses that the garment has 

been reapplied on the limb. The pump goes 

back to showing accumulated therapy time and 

continues counting from where it left off.

3.  Alternatively, pause and then restart the pump.