Compliance Monitoring; records and displays 

actual therapy as well as non-compliant time. Alerts 

upon garment removal to facilitate tracking and 

documentation of IPC therapy.

Enable/Disable Compliance 

Monitoring

1. 

Configure Compliance Monitoring: 

•  Make sure no garments are connected to the 

pump.

•  Press and hold the middle control button to 

access the setup screen.

2. 

Press the left button to cycle through the options: 

•  ON with Audible alert enabled
•  ON with Audible alert disabled
•  OFF (Pump run-hour time is displayed instead 

of accumulated therapy time)

ON/OFF

3.  Press the middle control button to select the 

displayed option.

Reset the Compliance Monitoring 

(accumulated therapy time) 

1.  When the pump is in Standby, press and hold 

the right control button for 2 seconds. The time 

flashes for 2 minutes. Press and hold the middle 

control button for 3 seconds to reset the time to 

zero.

2. 

Press the right control button again to confirm. If 

the button is not pressed within two minutes, the 

timer will NOT be reset and will go back to show 

the original data.

Start therapy

1.  Apply the garment to the patient.

2.  Connect the garment to the pump.

3.  Press the middle control button to start therapy.

“Click”

08.FL.01_01EN • 09/2022 • Page 1/2

QUICK REFERENCE GUIDE - USE

Flowtron ACS900 

Compliance Monitoring