background image

 

 

 

 

PROMETRA® II PROGRAMMABLE PUMP

   

 

 

 

 

 

Page 19 of 38

 

Precautions 

General 

 

Carefully read all instructions prior to use.  Follow all instructions. 

 

Certain equipment may cause electrical noise, which may interfere with programming.  If 

suspected, move the patient from the suspected source of interference to facilitate the 

programming procedure. Examples of equipment that may cause interference include cathode ray 

tube (CRT) monitors and large electric motors.   

 

Do not use accessories that are not referenced in these instructions for use.  Only use devices and 

accessories that are referenced for use with the Prometra® II Programmable Pump in these 

instructions.   

 

Safety and effectiveness for use in pediatric patients under 22 years old has not been investigated 

or established.  

 

The effects of implanting this device in patients with other implanted medical devices, other than 

neurostimulators, are unknown.   

 

Pain on injection that was not noted during previous injections may be an early sign of infection. 

 

Implant 

 

Implantation of this device and subsequent use, reprogramming, and refill should only be 

conducted by qualified medical personnel specifically trained for surgical implantation, use, and 

maintenance of the device.  Use of this device by non-qualified or untrained personnel could lead 

to serious consequences involving under- or over-dosage of Infumorph.  In the event of over-

dosage, refer to the approved Infumorph labeling for appropriate treatment. 

 

The pump and catheter system should be implanted carefully to avoid any sharp or acute angles, 

which could compromise the patency of the catheter lumen. 

 

Over-pressurization can damage the catheter.  Small syringes can generate very high pressures and 

may damage the catheter or catheter connection.  Do not use a syringe smaller than 10 mL when 

accessing the catheter access chamber. 

 

If therapy is discontinued for an extended period, the pump should be emptied of Infumorph and 

filled with a preservative-free 0.9% sterile saline solution and programmed to a low infusion rate to 

maintain catheter patency. 

 

Device Compatibility 

 

Pump accessories. 

Only use the Prometra II Programmable Pump with the accessories listed in 

these instructions for use.  Use of alternate accessories may result in damage to Prometra II 

components, less than adequate therapy, or increased risks to the patient. 

 

Pump.

 Only use with Prometra Programmer.  

 

Alcohol.

 Do not use alcohol on any part of the pump or catheter system. Alcohol is neurotoxic. 

 

Contrast media.

 Do not inject contrast media into the refill reservoir since this may damage the 

pump or impair pump function. 

 

External devices. 

Do not connect any external devices or pumps to the Prometra II Pump.  

Pressures generated by an external pump could damage the implanted pump/catheter system and 

result in serious patient injury or death. 

 

Therapeutic ultrasonics or lithotripsy - 

Use of therapeutic ultrasonic devices, such as 

Содержание 13827

Страница 1: ...PROMETRA II PROGRAMMABLE PUMP REF 13827 For use with Intrathecal Catheter MR Conditional Caution Federal Law USA restricts this device to sale by or on the order of a physician...

Страница 2: ...Programmable Features 23 Programming Medication Regimens 24 Pre Programmed Pump Settings 26 Pump Alarms 27 Implantation Instructions 28 Pre Implant Pump Programming Set Up 28 Pump Priming Preparation...

Страница 3: ...refer to the use of Infumorph Contents The following components are sterile and non pyrogenic 1 Prometra II Programmable Pump 1 Needle Non Coring 0 7 mm 22G x 38 mm 1 5 in 1 Needle Catheter Access 0 9...

Страница 4: ...of 38 The triple redundancy flow control system is designed to provide a precise and accurate flow rate The flow rate accuracy is independent of normal operating environmental conditions such as altit...

Страница 5: ...anium Usable Capacity 20 mL Precision Dosing System Material Titanium MP35N alloy Stainless steel Silicone rubber Refill Septum Septum material Silicone rubber Access needle Huber point 22G non coring...

Страница 6: ...natomy is insufficient to accommodate the size of the implanted pump or catheter The pump cannot be implanted 2 5 cm 1 in or less from the surface of the skin Deeper implants could interfere with sept...

Страница 7: ...rph labeling for appropriate treatment Clinicians implanting programming accessing or maintaining implanted programmable pumps must comply with the instructions for use Technical errors may result in...

Страница 8: ...Reservoir and Catheter into the patient This could result in drug overdose that could lead to serious patient injury or death Prior to initiating the MRI procedure the physician must determine if the...

Страница 9: ...agnetically induced deflection force and very strong torque The static and gradient magnetic fields produced by an MRI scanner could potentially interact with the pump and cause vibration However when...

Страница 10: ...iate discharge of the contents of the Drug Reservoir and Catheter into the patient This could result in drug overdose that could lead to serious patient injury or death The physician must determine if...

Страница 11: ...g Set the flow mode to a constant flow rate of 0 0 mg day Re inquire the pump and print inquiry page to confirm a constant flow rate of 0 0 mg day 3 Empty DrugReservoir Follow the procedures for empty...

Страница 12: ...ce needle through center Refill Port Septum until needle tip resides completely inside the Drug Reservoir c Pull a vacuum with the syringe for approximately 10 to 30 seconds to confirm Inlet Outlet Va...

Страница 13: ...igns and symptoms of drug overdose and develop a plan for immediate monitoring in a medically supervised and adequately equipped environment Resuscitative equipment should be available as should medic...

Страница 14: ...he drug was not removed from the Drug Reservoir prior to the MRI Following an MRI the FAVwill be closed and will prevent further drug delivery to occur until the pump is manually reset after the compl...

Страница 15: ...Errors d If no pump errors are displayed proceed to Step 3 FAV Reset Procedure 2 Clear Pump Errors a If pump errors are displayed from the Inquiry performed in Step 1 perform an Emergency Pump Stop u...

Страница 16: ...ge for approximately 10 to 30 seconds to confirm Inlet Outlet Valve closure Warning If any significant volume 1ml is retrieved it may be indicative that the pump Inlet Outlet Valves are open providing...

Страница 17: ...ontact patient s pump management physician The patient s medical records indicate the pump model and serial number implanted Flowonix provides medical chart labels to facilitate patient record documen...

Страница 18: ...Prometra II Pump Patient ID Card Card Front Card Back Prometra Pump Patient ID Card Card Front Card Back Prometra Pump X ray Prometra II Pump X ray Catheter Access Port Flow Activated Valve Pump Mode...

Страница 19: ...ph In the event of over dosage refer to the approved Infumorph labeling for appropriate treatment The pump and catheter system should be implanted carefully to avoid any sharp or acute angles which co...

Страница 20: ...erapies may have effects on pump operation that are not immediately apparent Potential Adverse Events The use of implanted pumps provides an important means of intrathecally delivering Infumorph Howev...

Страница 21: ...rare instances the development of an inflammatory mass at the tip of the implanted catheter may occur which can result in serious neurological impairment Patients should be monitored carefully at eac...

Страница 22: ...uffering from cancer pain requiring strong opioids chronic non malignant pain or required an implantable pump system replacement due to malfunction or battery depletion The average patient age at impl...

Страница 23: ...Site Irritation Implant Site Necrosis Edema Peripheral Hypersensitivity Extradural Abscess Implant Site Cellulitis Spinal Infection Viral Excoriation Hip Fracture2 Procedural Nausea Balance Disorder...

Страница 24: ...g Implantation of the Prometra II Programmable Pump System and subsequent use reprogramming and refill should only be conducted by qualified medical personnel specifically trained for surgical implant...

Страница 25: ...ose is delivered and the interval at which the dose is repeated are programmed Periodic Flow Regimen Demand Bolus The Demand Bolus regimen temporarily replaces the current dose regimen to deliver an i...

Страница 26: ...specific Drug Drug contained in pump No specified by user Conc Concentration of drug in pump Yes preset to 1 000 mg mL Accum Accumulator Volume Constant e g 2 010 L Yes pump specific Reservoir Volume...

Страница 27: ...setting the low reservoir volume and enabling the alarm refer to the Prometra Programmer Instructions Critical Error Alarm The Critical Error Alarm alerts patients and clinicians that the pump has st...

Страница 28: ...pump is still in the packaging and facing upward turn the programmer on press inquire and place the programmer over the pump 4 From the Main Menu select Setup Pump Setup and then Patient Name 5 Enter...

Страница 29: ...duction in concentration Expected concentration of drug in pump reservoir based on fill method Filling without rinsing Rinsing with 20 mL of drug 86 97 5 If rinsing the pump before filling rinse and d...

Страница 30: ...eted or if not completed canceled prior to attaching pump to catheter to avoid medication advancement into the catheter Do not cancel prior to 11 minutes to ensure fluid pathway is fully primed 6 If f...

Страница 31: ...theter as per the Prometra Programmable Infusions Systems Intrathecal Catheter IFU Implantation of the Prometra II Programmable Pump 1 USE STERILE TECHNIQUE Always inspect and aseptically prepare the...

Страница 32: ...king Warning Always save trimmed portion of catheter to measure length and calculate implanted catheter volume This calculation is required to prevent under or over medication 7 Slide catheter lock on...

Страница 33: ...This measurement is required to determine the volume of the implanted catheter 18 Calculate and record the implanted catheter length and volume Implanted Catheter Length cm 110 cm Trimmed Catheter Len...

Страница 34: ...changes such as skiing flying hot tubbing or saunas will not affect the operation of the pump Activities that involve an increase in environmental pressure of approximately 1 atmosphere or greater su...

Страница 35: ...changes such as skiing flying hot tubbing or saunas will not affect the operation of the pump Temperature related therapies such as deep heat therapy e g diathermy will not affect the operation of the...

Страница 36: ...of the pump This margin of safety was designed to offer your patients an additional measure of comfort and safety Pump flow rate accuracy was evaluated at multiple infusion rates 0 05mL day 28 8mL day...

Страница 37: ...rate in order to conserve energy required for pump operation The life of the pump is a minimum of 10 years at a drug delivery rate of 0 25mL day Drug Stability Drug stability has been tested for the...

Страница 38: ...act Flowonix Medical Inc to see if additional product information is available Trademarks are the property of their respective owners US and Foreign patents issued and pending Please consult www flowo...

Отзывы: