Configuration Settings            

 

 

 

 

 

 

Mounting the EchoPod

®

 

The EchoPod

®

 should always be mounted perpendicular to the liquid surface and 

installed using the provided Viton

®

 mounting gasket. Always use fittings, with thin wall 

mounting structures that isolate the transducer. This will provide the best performance 

over the lifetime of the product. 

The preferred mounting fitting for the EchoPod

®

 DL24 is the Reducer Bushing LM52-

1400 (2” thread x 1” thread). Further mounting products and solutions can be found 

on the Flowline website at 

www.flowline.com/fittings.php

. An instructional video on 

mounting ultrasonic level measurement technology is also available on the website. 

For installations in existing 2” fittings: 

Use LM52-1400 2” thread x 1” thread adapter (An adapter with 

an air gap around the 1 inch threads as shown is 

recommended) 

For installations in plastic tanks: 

1)  Use a Bulkhead fitting LM52-1890 1” inch Bulkhead 

fitting, 

2)  Use a Bulkhead fitting LM52-2890 in combination with 

LM52-1400 or, 

3)  Weld a plastic 1” half coupler to tank top. 

For installations in metal tanks: 

1)  Use the recommended bulkhead fittings as shown 

above or a LM52-1850 1” Flange. The flange fitting 

must have a riser for the threaded section. Drilling and 

tapping a blind flange is not recommended.   

2)  While installations directly into a 1”metal fitting are not 

recommended, acceptable results may be obtained if 

 the 1” fitting is a half coupling in  form and the outer 

 diameter of the coupler is tightly wrapped in vinyl tape 

 to dampen vibrations.  

 

For installations in open tanks and sumps: 

1)  Use Flowline's LM50-1001-1 side mount bracket, 

which includes a 2" x 1" thread reducer bushing. 

 

Using the drop-down menus on the left of the WebCal screen, set the configuration 

for your application requirements. When a selection does not apply to your 

application, “Not Applicable” will appear in the drop-down. Make sure all drop-downs 
are set appropriately for your application before moving to the Tank Level section. 

 

 

NOTE

: If you would like to start over, click the                                                                           

Clear Screen

 button on the right. 

 

 

 

 

 

 

Enter the appropriate tank level set points for your application. 

 

 

 

 

 

 

 

 

 

Tank Level Settings 

 

Units

. Display measurements in inches 

or centimeters. 

 

Sensor Height

. Distance measured from 

the bottom of the empty tank to the 

bottom of the transducer. Under factory 

configuration, this becomes the 4 mA set 

point. 

 

Fill Height

. Distance measured from the 

bottom of the empty tank to the maximum 

fill height within the tank. Under factory 

configuration, this becomes the 20 mA 

set point. 

 

Relay Set Points

. Distance measured 

from the bottom of the empty tank to each 

relay set point. 

Spec Tech Industrial Electric

               www.spectechind.com

888-773-2832

Spec Tech Industrial Electric

               www.spectechind.com

888-773-2832

Содержание EchoPod DL24 Series

Страница 1: ...ligent use 3 have been modified or altered 4 anyone other than service personnel authorized by Flowline have attempted to repair 5 have been involved in accidents or natural disasters or 6 are damaged during return shipment to Flowline Flowline reserves the right to unilaterally waive this warranty and dispose of any product returned to Flowline where 1 there is evidence of a potentially hazardous...

Страница 2: ...stalled Wiring should always be completed by a licensed electrician Supply voltage should never exceed 28 VDC Always use stepper relays between the sensor and external loads For DC circuits use a catch diode such as 1N4148 as shown above Protect the sensor from excessive electrical spikes by isolating the power whenever possible The sensor materials must be chemically compatible with the liquids t...

Страница 3: ...ration of your EchoPod should be performed prior to mounting since it requires connection to your PC Step 1 Install the WebCal Software Download WebCal from www flowline com onto a PC with the following minimum specifications Windows 2000 XP Vista 7 10 MB storage space 256 MB RAM 1 USB 2 0 port Double click the WebCal icon to install before proceeding to Step 2 You must have an active Internet con...

Страница 4: ...y vapor Select a 3 pipe a 2 pipe is usable but the minimum and attach the EchoPod with a coupling and reducer bushing The pipe length should run the measurement span and the bottom of the pipe should remain submerged at all times to prevent foam from entering the pipe Cut the bottom end of the pipe at 45 and drill a 1 4 pressure equalization hole high in the dead band The pumps should not drive li...

Страница 5: ...ot recommended 2 While installations directly into a 1 metal fitting are not recommended acceptable results may be obtained if the 1 fitting is a half coupling in form and the outer diameter of the coupler is tightly wrapped in vinyl tape to dampen vibrations For installations in open tanks and sumps 1 Use Flowline s LM50 1001 1 side mount bracket which includes a 2 x 1 thread reducer bushing Usin...

Страница 6: ...ng the EchoPod For updates to WebCal or to make sure you re using the most up to date version available click the Update tab on the top of the WebCal screen Write to Unit Send your configuration into EchoPod Wiring Diagram Open PDF wiring schematic of your configuration Advanced Configure advanced settings Read the section below on advanced setting options Factory Config Return EchoPod to the orig...

Отзывы: