The FloPatch FP120 is intended for use in the electromagnetic environment specified
below. The customer or the user of the FloPatch FP120 should assure that it is used in
such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level Electromagnetic Environment
–
Guidance
Conducted RF IEC
61000-4-6
3 Vrms
150 kHz to 80 MHz
6 Vrms
ISM bands inside
150 kHz to 80 MHz
N/A
Portable and mobile RF communications
equipment should be used no closer to any
part of the FloPatch FP120 including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d
= 1.2√
P
Radiated RF IEC
61000-4-3
3 V/m
80 MHz to 2,7 GHz
RF communication
equipment inside
80 MHz to 6 GHz
3 V/m
80 MHz to 2,7 GHz
RF communication
equipment inside
80 MHz to 6 GHz
d
= 1.2
√
P
80 MHz to 800 MHz
d
= 2.3
√
P
800 MHz to 2.7 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according
to
the
transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
should be less than the compliance level in each
frequency range
b
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the FloPatch FP120 is used
exceeds the applicable RF compliance level above, the FloPatch FP120 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the FloPatch FP120
b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.